Nimbus 4 and Nimbus Professional Mattresses
Reason for Recall / Hazard
Nimbus 4 and Nimbus Professional mattresses may over-inflate, due to an inner mattress component failure. This cannot be detected until the product is disassembled.
Product Description & Identification
Reason / Defects: Nimbus 4 and Nimbus Professional mattresses may over-inflate, due to an inner mattress component failure. This cannot be detected until the product is disassembled. Hazards: If the mattress over-inflates, patients could fall from the mattress and this could result in serious injury. What to do: Affected consumers will be contacted in writing. Consumers should follow the instructions provided in the customer letter and the instructional video, to puncture the grommet membrane so that excess air may escape. Removing the membrane will not affect the mattress functionality in any way. For more information, please contact Arjo Australia by phone on 1800 072 040.
Affected Products
Nimbus 4 and Nimbus Professional Mattresses Serial numbers: 1308000048 to 2000027834 ARTG 289068 (Arjo Australia Pty Ltd - Pressure alleviation pad)
Additional Source Details
| Field | Value |
|---|---|
| Guid › #text | f74d2367-be2c-468c-b360-9b869e764737 |
| Guid › @ is Perma Link | false |
| Link | https://www.productsafety.gov.au/search-consumer-product-recalls/nimbus-4-and-nimbus-professional-mattresses |
| Pub Date | Sun, 23 May 2021 00:00:00 +0000 |