Published: March 28, 2021Recall ID: 6d3a2833-0ba8-4c86-90dd-bdf55cce3762Category: generalCountry: Australia

Reason for Recall / Hazard

The rear legs of the walker have a small cylindrical mounting feature called a boss, which the castors are screwed into. If the welding is not sufficient between the leg and the boss, the boss can come loose and the castors will detach from the rear legs of the chassis.

Product Description & Identification

Reason / Defects: The rear legs of the walker have a small cylindrical mounting feature called a boss, which the castors are screwed into. If the welding is not sufficient between the leg and the boss, the boss can come loose and the castors will detach from the rear legs of the chassis. Hazards: If the weld fails, the boss comes loose and the castors detach from the legs of the walker, this could result in an injury and/or a fall. What to do: Sunrise Medical will contact affected consumers in writing. Consumers should send a completed acknowledgement form to Sunrise Medical to arrange for replacement legs to be fitted by an approved technician, within 4-6 months. Consumers can continue to use their MyWay chassis after conducting a pre-use inspection to ensure that the rear castor boss is securely fixed, all castors are moving freely and all brakes and resistance features are working correctly. For further information, contact Sunrise Medical Customer Service on 02 9678 6600.

Affected Products

MyWay Walker Chassis Product code: 173-1600, 173-2600 and 173-3600 Serial numbers range from LEC/220568 – LEC/272111 ARTG 298435 (Sunrise Medical Pty Ltd - Holder, patient)

Additional Source Details

FieldValue
Guid › #text6d3a2833-0ba8-4c86-90dd-bdf55cce3762
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/myway-walker-chassis
Pub DateSun, 28 Mar 2021 00:00:00 +0000
Extracted imagehttps://www.productsafety.gov.au/system/files/styles/large/private/MyWay%20Walker.jpg?itok=Nnw6kGY3

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert