Flocare Infinity II Enteral Feeding Pumps

Published: April 15, 2021Recall ID: 66c38698-8553-40b2-a674-93120f03a0deCategory: generalCountry: Australia

Reason for Recall / Hazard

In a situation where the device is not plugged into the main power supply and when running on battery power only, the device may unexpectedly switch off without sounding the alarm to indicate a low battery charge. This issue only affects Flocare Infinity Enteral Feeding Pumps. It does not affect any other Nutricia products.

Product Description & Identification

Reason / Defects: In a situation where the device is not plugged into the main power supply and when running on battery power only, the device may unexpectedly switch off without sounding the alarm to indicate a low battery charge. This issue only affects Flocare Infinity Enteral Feeding Pumps. It does not affect any other Nutricia products. Hazards: If the device switches off, this could cause a delay in the feeding routine and potentially lead to dehydration and diabetic complications. What to do: Nutricia Australia Pty Ltd will contact affected consumers in writing and advise consumers to follow the instructions in the customer letter: 1. Where possible, plug the device into the mains power during feeding therapy (e.g. during overnight feeding). 2. Fully charge the battery before operating the device without mains power (in a mobile use situation). The battery will be completely charged after approximately 6 hours. 3. If users notice that the battery must be charged more frequently than normal, please return the device to the service centre. 4. If the device is running on battery power only and users identify a significant decline in battery capacity or the device is not able to run a normal programmed therapy on a fully charged battery without hearing a low battery charge alarm, please return the device to the service centre. For further information, contact the Nutricia at Home team by phone on 1800 884 367 or the Nutricia Clinical Care Line on 1800 060 051.

Affected Products

Flocare Infinity II Enteral Feeding Pumps Product codes: 40407 and 40408 All serial numbers ARTG 119667 (Nutricia Australia Pty Ltd - Infusion pump, enteral feeding)

Additional Source Details

FieldValue
Guid › #text66c38698-8553-40b2-a674-93120f03a0de
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/flocare-infinity-ii-enteral-feeding-pumps
Pub DateThu, 15 Apr 2021 00:00:00 +0000
Extracted imagehttps://www.productsafety.gov.au/system/files/styles/large/private/Nutricare%20Flocare%20Infinity%20II%20Enteral%20Feeding%20Pump.jpg?itok=ZMtrvpAe

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert