Reason for Recall / Hazard

The affected masks contain magnets which can potentially affect the functioning of implanted medical devices such as metal splinters in the eyes, brain shunts, aneurysm clips, pacemakers, implantable cardioverter defibrillators, defibrillators, VP shunts, and neurostimulators used in and around the neck or any implanted medical device impacted by magnetic fields. Philips Respironics is updating the existing contraindications and warnings on the Instructions For Use (IFU) to strengthen the current labelling and warnings related to magnets.

Product Description & Identification

Reason / Defects: The affected masks contain magnets which can potentially affect the functioning of implanted medical devices such as metal splinters in the eyes, brain shunts, aneurysm clips, pacemakers, implantable cardioverter defibrillators, defibrillators, VP shunts, and neurostimulators used in and around the neck or any implanted medical device impacted by magnetic fields. Philips Respironics is updating the existing contraindications and warnings on the Instructions For Use (IFU) to strengthen the current labelling and warnings related to magnets. Hazards: If the mask magnets are placed less than 6 inches (approx. 15.24 cm) away from a metallic implant or device, the magnets may cause the device to not function as intended. This can result in serious injuries. Possible impacts may include seizure, arrhythmia, irregular blood pressure, change in heartbeats and cognitive issues. What to do: Consumers should read and follow the updated warnings and Instructions For Use (IFU) and take appropriate action: Patients, Household Members, Caregivers and Bed Partners with a medical implant and/or device(s): Must ensure the mask is kept at least 6 inches (approximately 15.24cm) away from the medical implant(s) and/or device(s) Patients should stop using the affected mask if the implant and/or medical device is contraindicated against the mask magnets. Patients should consult their physician immediately to determine if another mask can be used for their therapy and in the interim, switch to a non-magnetic mask if available, for continued therapy. Patients should properly dispose of the mask that has magnets after an alternative is obtained. If patients do not have implanted medical devices, or metallic splinters in their eyes, then no action related to patients is needed. If you are affected or need further information, please contact the Philips Sleep and Respiratory Care hotline on 180 830 517 to organise a suitable remediation which may include a replacement mask or clips to replace the magnets. For more details see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01171-1

Affected Products

Patient Interface devices (face and nasal masks) DreamWear Full Face Mask, DreamWisp Nasal Mask, Wisp and Wisp Youth Nasal Mask and Amara View Full Face Mask Multiple part numbers (see attached list) ARTGs 301663, 194980, 209222, 225674

Additional Source Details

FieldValue
Guid › #text718aca28-ae2a-4234-b1e3-312c7a8bf2ab
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/dreamwear-full-face-mask-dreamwisp-nasal-mask-with-over-the-nose-cushion-wisp-nasal-masks-and-wisp-youth-nasal-masks-amara-view-minimal-contact-full-face-mask-and-therapy-mask-3100-ncsp
Pub DateWed, 23 Nov 2022 00:00:00 +0000
Extracted imagehttps://www.productsafety.gov.au/system/files/styles/large/private/DreamWear%20Full%20Face%20Mask%20and%20DreamWisp%20Nasal%20Mask%20Wisp%20and%20Wisp%20Youth%20Nasal%20Mask%20and%20Amara%20View%20Full%20Face%20Mask.jpg?itok=VZ8NXM6v

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert