All MiniMed 640G, 670G, 700, 770G, and 780G Insulin Pumps

Published: June 9, 2022Recall ID: 2c86c2a3-93fa-4dd6-bcb1-12f1ff779780Category: generalCountry: Australia

Reason for Recall / Hazard

The gold battery cap contact may fall off when the user is charging the battery. If the contact falls off, the 'Insert battery' alarm will not clear automatically when the user inserts a new AA battery as the circuit is not complete. This will result in the alarm escalating to a siren and then the pump will shut down and will not power up.

Product Description & Identification

Reason / Defects: The gold battery cap contact may fall off when the user is charging the battery. If the contact falls off, the 'Insert battery' alarm will not clear automatically when the user inserts a new AA battery as the circuit is not complete. This will result in the alarm escalating to a siren and then the pump will shut down and will not power up. Hazards: If insulin delivery is suspended, this may lead to major hyperglycemia which can result in serious health complications. In severe cases of hyperglycemia, the user may experience diabetic ketoacidosis and may require hospitalisation or intervention from a medical professional. What to do: Medtronic will contact affected users in writing to advise them of the issue. Consumers are advised to follow the interim actions outlined in the customer letter and to have an alternate means of insulin therapy available. For more details see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-00725-1 For further information, please contact Medtronic Technical Support by phone on 1800 777 808.

Affected Products

All MiniMed 640G, 670G, 700, 770G, and 780G Insulin Pumps MiniMed pumps are intended for continuous insulin delivery and delivery of user activated insulin boluses, for the management of diabetes mellitus in persons requiring insulin. ARTGs 95763, 308140, 332201 and 376091

Additional Source Details

FieldValue
Guid › #text2c86c2a3-93fa-4dd6-bcb1-12f1ff779780
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/all-minimed-640g-670g-700-770g-and-780g-insulin-pumps
Pub DateThu, 09 Jun 2022 00:00:00 +0000

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert