700HPPK / 700HPM, High Pressure Pneumatic Calibration manifold, 21 MPA
Reason for Recall / Hazard
The plunger nut threads on the calibration manifold may fail, causing the plunger assembly to be ejected from the manifold at high speed.
Product Description & Identification
Reason / Defects: The plunger nut threads on the calibration manifold may fail, causing the plunger assembly to be ejected from the manifold at high speed. Hazards: There is a risk of serious injury to the user if they are struck by the ejected plunger assembly. What to do: Follow this link to register your recalled manifold on the Fluke website. Please include your full contact information, including an email and shipping address. Once registered, you will receive a return postage label so that it can be returned to FLUKE for repair and returned to you free of charge. For more information, contact Fluke on 1300 135 853 (Monday to Friday, 8:30 am – 5:00 pm AEST) or by emailing ordersanz@fluke.com
Affected Products
High-pressure pneumatic calibration manifold units included in 700HPPK test kits. Affected models: 700HPPK-NPT – Fluke No. 4784838 700HPPK-BSP – Fluke No. 4810462 700HPPK-MET – 4810443 All units with serial numbers below 6516999 are affected by this recall. Units with serial number 6517000 and above are not affected by this recall.
Additional Source Details
| Field | Value |
|---|---|
| Guid › #text | 08db20ad-f520-4ea8-a885-f8105f9149af |
| Guid › @ is Perma Link | false |
| Link | https://www.productsafety.gov.au/recalls/700hppk-700hpm-high-pressure-pneumatic-calibration-manifold-21-mpa |
| Pub Date | Wed, 31 Jul 2024 00:00:00 +0000 |
| Extracted image | https://www.productsafety.gov.au/system/files/styles/large/private/Serial%20Number.png?itok=AhorgXrs |