Home/Recalls/FDA-Z-1592-2026
FDA DevicesClass II

Volar Radius Plate with Angular Stability with the following description: Radius...

Published: April 1, 2026Recall ID: Z-1592-2026Category: devicesCountry: US

Reason for Recall / Hazard

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Product Description & Identification

Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.

Additional Source Details

FieldValue
CityLasnitzhohe
Event id98341
Address 1Autal 28
Code infoArticle Number (Lot Numbers): 21031-3 (679/042207, 679/032240, 679/0323115, 679/052317, 26/F70606, 679/032253, 643/032501, 643/072503, 643/102501), 21031-5 (43/0015, 643/102201, 643/102201, 679/0323121, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 643/102201, 643/102201, 643/102201, 643/102201, 679/0323120, 679/0323120), 21032-11 (679/112312), 21032-3 (679/032241, 643/0623, 679/042293, 679/0323122, 679/062367, 679/052318, 643/022501, 643/062501, 643/072501, 643/102502), 21032-5 (679/122136, 679/0323143, 679/042337), 21033-3 (679/062212, 679/022301, 679/042338, 679/022302), 21033-5 (643/0629, 26/F24901, 679/012311), 21034-3 (679/062221, 679/012312, 679/072217, 679/042339), 21034-5 (643/102202, 679/012314, 679/012313), 21035-3 (679/062222, 679/042308, 643/012501, 643/0548), 21036-3 (679/062213, 679/042309, 679/112317),
Postal codeN/A
Report date20260401
Product typeDevices
Product quantity220 units
Reason for recallUpdated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260123
Initial firm notificationE-Mail
Center classification date20260320

Overview

  • Recalling FirmI.T.S. GmbH
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Official Agency Alert