Home/Recalls/FDA-Z-1942-2026
FDA DevicesClass I

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Published: May 13, 2026Recall ID: Z-1942-2026Category: devicesCountry: US

Reason for Recall / Hazard

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

Product Description & Identification

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Additional Source Details

FieldValue
CityBothell
StateWA
Event id98554
Address 122002 26th Ave Se
Code infoUDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532
Postal code98021-4903
Report date20260513
Product typeDevices
Product quantity42
Reason for recallRespiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260325
Center classification date20260501

Overview

  • Recalling FirmVentec Life Systems, Inc.
  • StatusOngoing
  • Risk LevelClass I
  • DistributionUS: OH, GA, CA, MO