Reason for Recall / Hazard

Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.

Product Description & Identification

The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips; URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein; Vivoo Ketone Test

Affected Products

The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips; URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein; Vivoo Ketone Test

Additional Source Details

FieldValue
CityChangchun
StateN/A
Event id97801
Address 12336, Tianwei Road
Address 2Beihu Science and Technology Development Zone
Code infoName/Lot(Expiration): The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips/2024032508(03/24/2026), 2024071602(07/15/2026), 2024102108(10/20/2026); URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein/2025043007(04/29/2027); Vivoo Ketone Test/2024013122(01/30/2026)
Postal codeN/A
Report date20260211
Product typeDevices
Product quantity32,500
Reason for recallTest strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251122
Initial firm notificationLetter
Center classification date20260204

Overview

  • Recalling FirmChangchun Wancheng Bio-Electron Co., Ltd.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution in the states of TX, GA, CA.