Home/Recalls/AU-179266af-5942-4517-bf9b-e0a5b24a1b9c
Australia ProductsPharmacy and me...

Tandem t:slim X2 Insulin Pump

Published: May 27, 2020Recall ID: 179266af-5942-4517-bf9b-e0a5b24a1b9cCategory: generalCountry: Australia

Reason for Recall / Hazard

The pump battery may deplete faster than expected in certain circumstances. Only patients using the pump displaying Dexcom G5 Continuous Glucose Monitor (CGM) values are affected by this issue. If a Transmitter ID is entered incorrectly into the pump, the pump will be unable to connect to the transmitter and will continue to search for that transmitter until the battery charge is depleted or a correct Transmitter ID is entered.

Product Description & Identification

Reason / Defects: The pump battery may deplete faster than expected in certain circumstances. Only patients using the pump displaying Dexcom G5 Continuous Glucose Monitor (CGM) values are affected by this issue. If a Transmitter ID is entered incorrectly into the pump, the pump will be unable to connect to the transmitter and will continue to search for that transmitter until the battery charge is depleted or a correct Transmitter ID is entered. Hazards: Depletion of the pump battery may lead to interruption of insulin delivery, this could result in the user developing dangerously high blood glucose levels (hyperglycaemia), which can lead to serious health complications. What to do: This issue will be rectified via a software update, scheduled for release in August 2020. AMSL Diabetes Customer Care will be individually contacting patients to arrange to complete the update of each insulin pump. Consumers are requested to adhere to the following precautions until this software update is made available: 1. Please take precautions when entering the Transmitter ID into the t:slim X2 insulin pump to ensure that the Transmitter ID is correct before starting a Sensor Session. 2. Periodically check the battery level indicator of the pump; 3. Charge the pump for a short period of time every day (10 to 15 minutes); 4. Avoid frequent full battery discharges; and 5. Ensure that the “Bluetooth Settings” in the “Pump Settings” menu is turned off. For further information, contact AMSL Diabetes Customer Care on 1300 851 056.

Affected Products

Tandem Diabetes Care t:slim X2 Insulin Pump The Tandem Diabetes Care t:slim X2 Insulin Pump is intended for continuous insulin delivery and delivery of user activated insulin boluses, for the management of diabetes mellitus in persons requiring insulin. ARTG 304681 (Australasian Medical & Scientific Ltd - Infusion pump, insulin, ambulatory)

Additional Source Details

FieldValue
Guid › #text179266af-5942-4517-bf9b-e0a5b24a1b9c
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/tandem-tslim-x2-insulin-pump
Pub DateWed, 27 May 2020 00:00:00 +0000
Extracted imagehttps://www.productsafety.gov.au/system/files/styles/large/private/Tslim%20X2%20Insulin%20Pump.jpg?itok=F5EIaK5p

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert