T34 Syringe Pump

Published: March 10, 2021Recall ID: 6cb90fcd-6180-4a37-9877-be481a8ba7deCategory: generalCountry: Australia

Reason for Recall / Hazard

The voltage of the 9V/6LR61 battery may drop prematurely or earlier than expected in T34 2nd and 3rd edition devices. This may cause an interruption to the infusion, following events of: 1. Premature 'low battery' or 'end battery' alarms 2. Unexpected pump shut down with or without alarm. Battery brands/lots that have lower and more stable internal resistance are not likely to cause significant voltage dips and therefore not likely to trigger early low/end battery alarms or pump shut down.

Product Description & Identification

Reason / Defects: The voltage of the 9V/6LR61 battery may drop prematurely or earlier than expected in T34 2nd and 3rd edition devices. This may cause an interruption to the infusion, following events of: 1. Premature 'low battery' or 'end battery' alarms 2. Unexpected pump shut down with or without alarm. Battery brands/lots that have lower and more stable internal resistance are not likely to cause significant voltage dips and therefore not likely to trigger early low/end battery alarms or pump shut down. Hazards: If there is an interruption to the infusion, or the pump shuts down unexpectedly, this could result in the patient receiving insufficient delivery of medication or fluids, which can lead to serious health complications. What to do: REM Systems will contact affected customers in writing. Consumers are requested to take the following actions for devices currently in clinical use: 1. Continue use of the device with 9V/6LR61 type batteries as per the Directions For Use (DFU) instructions; 2. As a precaution, always have a spare set of high quality batteries; 3. Be mindful that battery life may be affected by factors such as temperature and battery manufacturer; 4. Maintain infusion monitoring and best clinical practices to detect and react to system alarms as per DFU instructions; and 5. Report without delay any device complaint including events of early/premature battery alarms, or any instances of unexpected pump shut down without alarm. For further information, contact REM Systems Customer Service by phone on 1800 737 222

Affected Products

T34 Syringe Pump All serial numbers ARTG 131232 (REM Systems Pty Ltd - Infusion pump, syringe) For the provision of small dose infusion therapy, the T34 Syringe Pump gradually administers small precise amounts of fluid and/or medications to a patient.

Additional Source Details

FieldValue
Guid › #text6cb90fcd-6180-4a37-9877-be481a8ba7de
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/t34-syringe-pump
Pub DateWed, 10 Mar 2021 00:00:00 +0000
Extracted imagehttps://www.productsafety.gov.au/system/files/styles/large/private/T34%20Syringe%20Pump.jpg?itok=aGKI5-O1

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert