Reason for Recall / Hazard

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Product Description & Identification

SMV BodyTrack, System, Tomography, Computed, Emission

Affected Products

SMV BodyTrack, System, Tomography, Computed, Emission

Additional Source Details

FieldValue
CityTirat Carmel
Event id98133
Address 14, Hayozma St
Code infoUDI/DI Not applicable: All serial numbers in distribution
Postal codeN/A
Report date20260121
Product typeDevices
Product quantity12 units
Reason for recallGE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251210
Initial firm notificationTelephone
Center classification date20260109

Overview

  • Recalling FirmGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
Official Agency Alert