Home/Recalls/CANADA-RA-76709
Health CanadaType II

REVOLUTION CT (2021-10-11)

Published: October 11, 2021Recall ID: RA-76709Category: generalCountry: Canada

Product Description & Identification

REVOLUTION CT (2021-10-11)

Additional Source Details

FieldValue
Panels › 1 › Text<b>Issue:</b> Medical Devices<BR/><b>Audience:</b> General Public, Healthcare Professionals, Hospitals<BR/><b>Category:</b> Medical Device<BR/><b>Alert type:</b> Medical Device Recall<BR/><b>Department:</b> Health Canada<BR/><b>Hazard classification:</b> Type II<BR/>
Panels › 1 › TitleSummary
Panels › 1 › Panel Namebasic_details
Panels › 2 › TextAffected Products<BR/><p> A. REVOLUTION CT - MAIN UNIT<br /> B. REVOLUTION CT ES<br /> C. REVOLUTION APEX</p>
Panels › 2 › TitleIntroduction
Panels › 2 › Panel Nameintro_text
Panels › 3 › Text There is a potential for the accumulated dose to be incorrectly displayed in certain situations. During an interventional exam using the Smartstep option, if the table height is adjusted after the exam starts, the Smartstep display will show a value that is higher than the actual dose delivered.
Panels › 3 › TitleReason
Panels › 3 › Panel Namecms_reason_reason_recall
Panels › 4 › Text<b>Brand name:</b> A. REVOLUTION CT - MAIN UNIT<br/><b>Lot or serial number:</b> 00000440051CN1<br /> 00000441839CN8<br /> 00000443342CN1<br /> 00000445060CN7<br /> REVVX1600054CN<br /> REVVX1700084CN<br /> REVVX1900034CN<br /> REVVX1900098CN<BR/><b>Model or catalog number:</b> 5590000<br /> 5590000-6<BR/><b>Companies:</b><br/><dl><dt>Manufacturer</dt><dd>GE Medical Systems LLC</dd><dd>3200 North Grandview Blvd</dd><dd>Waukesha</dd><dd>53188</dd><dd>UNITED STATES</dd></dl>
Panels › 4 › TitleA. REVOLUTION CT - MAIN UNIT
Panels › 4 › Panel Nameproduct_1
Panels › 5 › Text<b>Brand name:</b> B. REVOLUTION CT ES<br/><b>Lot or serial number:</b> REVV81900029CN<br /> REVV81900035CN<BR/><b>Model or catalog number:</b> 5590000-15<BR/><b>Companies:</b><br/><dl><dt>Manufacturer</dt><dd>GE Medical Systems LLC</dd><dd>3200 North Grandview Blvd</dd><dd>Waukesha</dd><dd>53188</dd><dd>UNITED STATES</dd></dl>
Panels › 5 › TitleB. REVOLUTION CT ES
Panels › 5 › Panel Nameproduct_2
Panels › 6 › Text<b>Brand name:</b> C. REVOLUTION APEX<br/><b>Lot or serial number:</b> REV2A2000017CN<br /> REV2A2000029CN<BR/><b>Model or catalog number:</b> 840682146616<BR/><b>Companies:</b><br/><dl><dt>Manufacturer</dt><dd>GE Medical Systems LLC</dd><dd>3200 North Grandview Blvd</dd><dd>Waukesha</dd><dd>53188</dd><dd>UNITED STATES</dd></dl>
Panels › 6 › TitleC. REVOLUTION APEX
Panels › 6 › Panel Nameproduct_3
Panels › 7 › Text<b>Identification number:</b> RA-76709<BR/>
Panels › 7 › TitleIdentification number
Panels › 7 › Panel Nameid_numbers
Recall IdRA-76709
Start date1633910400
Date published1634860800

Overview

  • Risk LevelType II
Official Agency Alert