Reason for Recall / Hazard

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Product Description & Identification

Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: The ¿-Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 ¿ 1.5 mL (4 Concentrations) CAL 1 0 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Affected Products

Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: The ¿-Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 ¿ 1.5 mL (4 Concentrations) CAL 1 0 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Additional Source Details

FieldValue
CityMalvern
StatePA
Event id98232
Address 1201 Great Valley Pkwy
Address 2N/A
Code infoModel/Catalog Number: 235454; UDI-DI: 04987270235454; Lot Numbers: D8C6022U, D8C6044U;
Postal code19355-1308
Report date20260211
Product typeDevices
Product quantity128 units
Reason for recallCustomers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251211
Initial firm notificationLetter
Center classification date20260205

Overview

  • Recalling FirmFujirebio Diagnostics, Inc.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Official Agency Alert