Home/Recalls/FDA-Z-1090-2026
FDA DevicesClass II

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PRO...

Published: January 21, 2026Recall ID: Z-1090-2026Category: devicesCountry: US

Reason for Recall / Hazard

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Product Description & Identification

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Affected Products

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Additional Source Details

FieldValue
CityNorthfield
StateIL
Event id98197
Address 13 Lakes Dr
Code infoUDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426
Postal code60093-2753
Report date20260121
Product typeDevices
Product quantity12 units
Reason for recallMedline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251216
Initial firm notificationLetter
Center classification date20260114

Overview

  • Recalling FirmMedline Industries, LP
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution.