MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit...

Published: January 21, 2026Recall ID: Z-1024-2026Category: devicesCountry: US

Reason for Recall / Hazard

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Product Description & Identification

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Additional Source Details

FieldValue
CityMarlborough
StateMA
Event id98260
Address 1111 Locke Dr
Code infoPN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Postal code01752-7236
Report date20260121
Product typeDevices
Product quantity33
Reason for recallPotential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Voluntary mandatedFDA Mandated
Recall initiation date20251210
Center classification date20260113

Overview

  • Recalling FirmREPLIGEN CORPORATION
  • StatusOngoing
  • Risk LevelClass II
  • DistributionU.S.
Official Agency Alert