Home/Recalls/AU-275d9437-1bbe-40bf-8073-3f9f1deca04b
Australia ProductsPharmacy and me...

Invacare Perfecto2 V Oxygen Concentrator

Published: January 19, 2022Recall ID: 275d9437-1bbe-40bf-8073-3f9f1deca04bCategory: generalCountry: Australia

Reason for Recall / Hazard

The sound abatement washer may break down resulting in metal on metal wear inside the Pressure Equalisation (PE) valve. This may lead to high temperatures and overheating, separation of the shroud from the device, and loud noise, and can result in a fire that will shut down the device.

Product Description & Identification

Reason / Defects: The sound abatement washer may break down resulting in metal on metal wear inside the Pressure Equalisation (PE) valve. This may lead to high temperatures and overheating, separation of the shroud from the device, and loud noise, and can result in a fire that will shut down the device. Hazards: If the device shuts down this can cause hypoxemia (low oxygen in the blood) from "loss of use". There is an increased risk of fire, which may result in serious injury and property damage. What to do: Consumers should check the model and serial number of their machine to see if their device is affected. Invacare Australia will contact affected users in writing with instructions on how to to make arrangements to replace the PE valve assembly. The replacement of the PE valve assembly will be done within 6 months. In the interim, consumers should continue to use the device in accordance with its labelling/user manual. Labelling that accompanies the device instructs the user to always have a backup source of oxygen readily available. For further information, consumers can contact Invacare Australia on 1800 460 460.

Affected Products

Invacare Perfecto2 V Oxygen Concentrator Product Model: IRC5PO2VAWL2 Serial numbers: 18BF007341 to 181F024339 ARTG: 162637 (Invacare Australia Pty Ltd - Oxygen concentrators)

Additional Source Details

FieldValue
Guid › #text275d9437-1bbe-40bf-8073-3f9f1deca04b
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/invacare-perfecto2-v-oxygen-concentrator
Pub DateWed, 19 Jan 2022 00:00:00 +0000

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert