Home/Recalls/AU-63e6d6a6-85c8-432b-870d-60ffbb1717e6
Australia ProductsPharmacy and me...

Inogen One G3 Concentrator

Published: February 27, 2020Recall ID: 63e6d6a6-85c8-432b-870d-60ffbb1717e6Category: generalCountry: Australia

Reason for Recall / Hazard

When connected to a fully charged 8-cell battery as its only power source, an operating condition may cause an abrupt shut down of the Inogen One G3 Portable Oxygen Concentrator (POC). If no action is taken, the POC may become unresponsive for up to one minute, after which it will behave as though a battery was just connected with a standard flash screen on display. The POC will then be ready for another startup. If powered on again, this cycle may need to be repeated until the 8-cell battery has been adequately discharged allowing the POC to operate as normal.

Product Description & Identification

Reason / Defects: When connected to a fully charged 8-cell battery as its only power source, an operating condition may cause an abrupt shut down of the Inogen One G3 Portable Oxygen Concentrator (POC). If no action is taken, the POC may become unresponsive for up to one minute, after which it will behave as though a battery was just connected with a standard flash screen on display. The POC will then be ready for another startup. If powered on again, this cycle may need to be repeated until the 8-cell battery has been adequately discharged allowing the POC to operate as normal. Hazards: Risk of interruption to the supply of oxygen to users, which may lead to health complications. What to do: Customers are advised that arrangements will be made for the products to be repaired and/or replaced. An Air Liquide Healthcare representative will contact customers via phone to initiate the replacement POC, and collect the affected concentrator. For further information, contact Air Liquide Healthcare Customer Service Centre on 1300 360 202.

Affected Products

Inogen One G3 Concentrator Portable lightweight medical oxygen machine Part Numbers: 19252750, 19252534, 19285707, 19285618, 19285718, 19285621, 19285334, 19252739, 19285727, 19284756, 19252747, 19285626, 19285728, 19252726, 19284685, 19282715, 19285619, 19252733 ARTG 225950 (Air Liquide Healthcare Pty Ltd - Stationary oxygen concentrator)

Additional Source Details

FieldValue
Guid › #text63e6d6a6-85c8-432b-870d-60ffbb1717e6
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/inogen-one-g3-concentrator
Pub DateThu, 27 Feb 2020 00:00:00 +0000

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert