Home/Recalls/FDA-Z-1809-2026
FDA DevicesClass II

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

Published: April 22, 2026Recall ID: Z-1809-2026Category: devicesCountry: US

Reason for Recall / Hazard

it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.

Product Description & Identification

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

Additional Source Details

FieldValue
CityIrvine
StateCA
Event id98528
Address 111 Hughes
Address 2N/A
Code infoILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.
Postal code92618-1902
Report date20260422
Product typeDevices
Product quantity15 units
Reason for recallit was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20250813
Initial firm notificationLetter
Center classification date20260410

Overview

  • Recalling FirmBeta Bionics, Inc.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.
Official Agency Alert