Heraeus, PALAMIX uno. Material Number: 66057893.

Published: May 6, 2026Recall ID: Z-1997-2026Category: devicesCountry: US

Reason for Recall / Hazard

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Product Description & Identification

Heraeus, PALAMIX uno. Material Number: 66057893.

Additional Source Details

FieldValue
CityWehrheim
Event id98608
Address 1Philipp-Reis-Str. 8
Code infoMaterial Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
Postal codeN/A
Report date20260506
Product typeDevices
Product quantity51,771 units
Reason for recallBased on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260325
Initial firm notificationLetter
Center classification date20260429

Overview

  • Recalling FirmHeraeus Medical GmbH (Dental Division)
  • StatusOngoing
  • Risk LevelClass III
  • DistributionWorldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;