DDU-2000 Series Automated External Defibrillator (AED)
Reason for Recall / Hazard
An electrical component may cause the product to abort a shock delivery, or reset unexpectedly.
Product Description & Identification
Reason / Defects: An electrical component may cause the product to abort a shock delivery, or reset unexpectedly. Hazards: An aborted shock or unexpected delivery may result in patient death under emergency situations. What to do: Device Technologies Australia will contact impacted customers to organise the return and replacement of affected units. Customers are advised that if alternate units are available for use, please return affected units to Device Technologies and they will be replaced. If no alternate units are available, please inform Device Technologies and continue using the device with caution until replacement is provided. For more information, contact Device Technologies Customer Service on 1300 338 423.
Affected Products
DDU-2000 Series Automated External Defibrillator (AED) DT Product Codes: DCF-E2310-EN and DCF-E2460EN Serial numbers: 400125196 400125197 400125203 400101350 400101365 ARTG 155076 (Device Technologies Australia Pty Ltd - Defibrillator, semi-automated)