Home/Recalls/FDA-Z-1185-2026
FDA DevicesClass II

Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 100...

Published: February 4, 2026Recall ID: Z-1185-2026Category: devicesCountry: US

Reason for Recall / Hazard

due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.

Product Description & Identification

Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters

Affected Products

Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters

Additional Source Details

FieldValue
CityMesa
StateAZ
Event id98030
Address 13110 N Oakland Ste 101
Address 2N/A
Code infoCatalog/Model Numbers/ Lot Numbers / UDI codes: 10001 14250616-1019106 (01) 0 0850016 97600 0 (17) 251030 (10) SHOWCASE 10011 14250616-1019107 (01) 0 0850016 97609 3 (17) 251030 (10) SHOWCASE 10040c 50250624-1019120 (01) 0 0850016 97636 9 (17) 251030 (10) SHOWCASE 10040 50250624-1019119 (01) 0 0850016 97635 2 (17) 251030 (10) SHOWCASE 10032 14250616-1019108 (01) 0 0850016 97631 1 (17) 251030 (10) SHOWCASE 10041 50250522-1018643 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10041 50250624-1019121 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10021 15F250603-1018834 (01) 0 0850016 97619 2 (17) 251030 (10) SHOWCASE
Postal code85215-1168
Report date20260204
Product typeDevices
Product quantity499 units
Reason for recalldue to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251212
Initial firm notificationLetter
Center classification date20260127

Overview

  • Recalling FirmNIPRO Technical Services, Inc.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionU.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Official Agency Alert