Claviculaplate with Angular Stability, with the below descriptions and associate...
Reason for Recall / Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Product Description & Identification
Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article Numbers: 21122-10, 21122-6, 21122-8. 4. Clavicle Plate, Lateral; Article Numbers: 21123-4, 21124-4, 21127-6, 21128-6.
Affected Products
Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article Numbers: 21122-10, 21122-6, 21122-8. 4. Clavicle Plate, Lateral; Article Numbers: 21123-4, 21124-4, 21127-6, 21128-6.
Additional Source Details
| Field | Value |
|---|---|
| City | Lasnitzhohe |
| Event id | 98341 |
| Address 1 | Autal 28 |
| Code info | 1. Article Number (Lot Numbers): 21116-8 (679/092230), 21117-10 (679/032313, 679/092231), 21118-10 (679/032314, 679/072047, 679/092232). 2. Article Number (Lot Numbers): 21121-6 (679/092236, 679/092329), 21121-8 (679/052226). 3. Article Number (Lot Numbers): 21122-10 (00/0425-02, 00/0825-02, 679/022443, 679/062310, 86/043372), 21122-6 (00/0325-01, 00/0425-01, 679/042340, 679/062334, 679/122269, 86/043366), 21122-8 (00/0325-08, 679/022243, 679/022304, 679/022451, 679/042209, 679/042214, 679/052319, 679/102368, 86/043368). 4. Article Number (Lot Numbers): 21123-4 (00/0810-06, 679/052014, 679/062216), 21124-4 (679/042333, 679/062217, 679/072346), 21127-6 (679/052228), 21128-6 (679/042341). |
| Postal code | N/A |
| Report date | 20260401 |
| Product type | Devices |
| Product quantity | 322 units |
| Reason for recall | Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260123 |
| Initial firm notification | |
| Center classification date | 20260320 |
Overview
- Recalling FirmI.T.S. GmbH
- StatusOngoing
- Risk LevelClass II
- DistributionUS Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.