CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Published: March 18, 2026Recall ID: Z-1553-2026Category: devicesCountry: US

Reason for Recall / Hazard

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment

Product Description & Identification

CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Additional Source Details

FieldValue
CitySunto-Gun
StateN/A
Event id98346
Address 1454-32
Address 2Higashino; Nagaizumi-Cho
Code infoREF/UDI-DI/ Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B96697/14987666543016/2025061442, 2025061446, 2025061452, 2025061454, 2025061464, 2025071479, 2025071488, 2025081513, 2025081516, 2025091536, 2025091538, 2025091543, 2025091559, 2025091561, 2025111598, 2025111602, 2025111604, 2025111605, 2025111606, 2025121618, 2025121622, 2025121623. B96698/14987666543023/2025061449, 2025061451, 2025061457, 2025061467, 2025071475, 2025071476, 2025071478, 2025071481, 2025071495, 2025071497, 2025081503, 2025081524, 2025091539, 2025091549, 2025091551. Older Instrument Replaced with PC preinstalled with Microsoft SQL Server 2022 Express: B23279/14987666541296/2014011316, 2014071647, 2014121858, 2014121906. B23280/14987666541302/2016022593. B23281/14987666541319/5094354. B96697/14987666543016/2019074683. B96698/14987666543023/2017063377, 2019114891, 2024101288
Postal codeN/A
Report date20260318
Product typeDevices
Product quantity47
Reason for recallA delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260130
Initial firm notificationLetter
Center classification date20260311

Overview

  • Recalling FirmBeckman Coulter Mishima K.K.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.