Home/Recalls/FDA-Z-1571-2026
FDA DevicesClass II

Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a de...

Published: March 25, 2026Recall ID: Z-1571-2026Category: devicesCountry: US

Reason for Recall / Hazard

There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.

Product Description & Identification

Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150

Additional Source Details

FieldValue
CityWaukesha
StateWI
Event id98428
Address 13000 N Grandview Blvd
Address 2N/A
Code infoProduct ID Numbers: 4446-2-Centricity Universal Viewer-00558643 100256-3-Centricity Universal Viewer-00558700 101044-1-Centricity Universal Viewer-00558694 13240-2-Centricity Universal Viewer-00558103 4441-1-Centricity Universal Viewer-00558677 4578-1-Centricity Universal Viewer-00558340 10831-2-Centricity Universal Viewer-00558682 656341-Centricity Universal Viewer-01431499 50567-Centricity Universal Viewer-02176369 2690973-Centricity Universal Viewer-01454622 EG1526-Centricity Universal Viewer-00158976 A91901-Centricity Universal Viewer-00044549 M3488569-Centricity Universal Viewer-00071707 25540620 - HC4105-Centricity Universal Viewer-00001393 416005-Centricity Universal Viewer-00260808 1868708X - 556162-Centricity Universal Viewer-00157857 HKOB01-Centricity Universal Viewer-01282268 BMH-Centricity Universal Viewer-01820534 JIPMER-Centricity Universal Viewer-01820555 4347967-Centricity Universal Viewer-01474540 651342-Centricity Universal Viewer-01365218 651342-Centricity Universal ... [TRUNCATED]
Postal code53188-1615
Report date20260325
Product typeDevices
Product quantity40 units
Reason for recallThere is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260130
Initial firm notificationLetter
Center classification date20260316

Overview

  • Recalling FirmGE Medical Systems, LLC
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide - US Nationwide distribution.
Official Agency Alert