CellMek SPS Sample Preparation System, REF: C44603, with software

Published: February 11, 2026Recall ID: Z-1235-2026Category: devicesCountry: US

Reason for Recall / Hazard

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Product Description & Identification

CellMek SPS Sample Preparation System, REF: C44603, with software

Additional Source Details

FieldValue
CityMiami
StateFL
Event id98207
Address 111800 Sw 147th Ave
Address 2N/A
Code infoUDI-DI: 15099590750312. Software Version 2.3.90.0
Postal code33196-2500
Report date20260211
Product typeDevices
Product quantity50
Reason for recallSample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251223
Initial firm notificationLetter
Center classification date20260202