Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Published: February 11, 2026Recall ID: Z-1300-2026Category: devicesCountry: US

Reason for Recall / Hazard

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.

Product Description & Identification

Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Affected Products

Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Additional Source Details

FieldValue
CityRastatt
StateN/A
Event id98159
Address 1Kehler Str. 31
Address 2N/A
Code infoModel Number: 701055720; UDI-DI: 04037691816432; Serial Numbers: All units impacted;
Postal codeN/A
Report date20260211
Product typeDevices
Product quantity3050 units
Reason for recallInternal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260109
Initial firm notificationLetter
Center classification date20260205

Overview

  • Recalling FirmMaquet Cardiopulmonary Gmbh
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
Official Agency Alert