Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: ...
Reason for Recall / Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Product Description & Identification
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
Affected Products
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
Additional Source Details
| Field | Value |
|---|---|
| City | Stockholm |
| Event id | 98117 |
| Address 1 | Eugeniavagen 18c |
| Code info | Lot Code: UDI: 0735000201054920220616, 0735000201067920221007, 0735000201073020230913 GTIN: 07350002010549, 07350002010679, 07350002010730 Serial Numbers: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Revisions: RayStation 12A, 12A SP 1, 12A SP2 Expiration Date: 2028-07-07 |
| Postal code | N/A |
| Report date | 20260128 |
| Product type | Devices |
| Product quantity | 8 |
| Reason for recall | Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20251128 |
| Initial firm notification | |
| Center classification date | 20260116 |
Overview
- Recalling FirmRAYSEARCH LABORATORIES AB
- StatusOngoing
- Risk LevelClass II
- DistributionRefer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS