Home/Recalls/AU-d18485d2-cb8f-42b1-ae36-f992fd924144
Australia ProductsIndoor furnitur...

Bock Nursing Care Beds - Battery

Published: February 12, 2020Recall ID: d18485d2-cb8f-42b1-ae36-f992fd924144Category: generalCountry: Australia

Reason for Recall / Hazard

In beds with double drives, manufactured in the period from January 2012 to May 2013, some batteries may contain diodes which may be defective.

Product Description & Identification

Reason / Defects: In beds with double drives, manufactured in the period from January 2012 to May 2013, some batteries may contain diodes which may be defective. Hazards: Failure of the diode may lead to damage of the emergency lowering battery. If the bed cannot be lowered this may lead to injury. What to do: Consumers with impacted stock should contact Alphacare to organise correction. Alphacare will arrange a technician to visit and make the necessary corrections. For further information, contact Alphacare Customer Service on 1300 930 930 or via the Alphacare website at https://alphacare.com.au/contact/

Affected Products

Bock Nursing Care Beds - Battery Model Number: 263 Serial Number: 91391986-000 ARTG 209294 (Alphacare Pty Ltd - Bed, general-purpose)

Additional Source Details

FieldValue
Guid › #textd18485d2-cb8f-42b1-ae36-f992fd924144
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/bock-nursing-care-beds-battery
Pub DateWed, 12 Feb 2020 00:00:00 +0000