Home/Recalls/FDA-Z-1587-2026
FDA DevicesClass III

Battery Charging Station; Model: 0998-00-0802;

Published: March 25, 2026Recall ID: Z-1587-2026Category: devicesCountry: US

Reason for Recall / Hazard

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.

Product Description & Identification

Battery Charging Station; Model: 0998-00-0802;

Affected Products

Battery Charging Station; Model: 0998-00-0802;

Additional Source Details

FieldValue
CityMahwah
StateNJ
Event id98432
Address 11300 Macarthur Blvd
Address 2N/A
Code infoModel: 0998-00-0802; UDI: 10607567111964; Serial Numbers: 200019347EAI, 200019847EAI, 200019447EAI, 210018625EAI, 210017925EAI, 210017525EAI, 210018425EAI, 210020144EAI, 210020944EAI, 210022544EAI, 210020244EAI, 210022144EAI, 210020044EAI, 210020544EAI, 210021344EAI, 210023244EAI, 210018025EAI, 210019025EAI, 210021144EAI, 210021044EAI, 210022344EAI, 210019225EAI, 210019125EAI, 210018525EAI, 210022844EAI, 200017647EAI, 210019725EAI, 210022744EAI;
Postal code07430-2052
Report date20260325
Product typeDevices
Product quantity28 units (20 US, 8 OUS)
Reason for recallThe Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260206
Initial firm notificationLetter
Center classification date20260319

Overview

  • Recalling FirmDatascope Corp.
  • StatusOngoing
  • Risk LevelClass III
  • DistributionWorldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.