Automated Impella Controller (AIC), labeled as the following with corresponding ...
Reason for Recall / Hazard
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Product Description & Identification
Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.
Affected Products
Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.
Additional Source Details
| Field | Value |
|---|---|
| City | Danvers |
| State | MA |
| Event id | 98393 |
| Address 1 | 22 Cherry Hill Dr |
| Code info | All products below with software version 8.2.2 through 8.5 are affected. 1. Product Code: 0042-0000-CA; GTIN: 00813502011272; Serial Numbers: IC2035, IC9416. 2. Product Code: 0042-0000-EU; GTIN: 00813502011289; Serial Numbers: IC1677, IC3049, IC3099, IC3102, IC4358, IC4329, IC5774, IC5757, IC6580, IC6581. 3. Product Code: 0042-0000-UK; GTIN: 00813502011296; Serial Numbers: IC2023, IC1910, IC2704, IC3325, IC3312. 4. Product Code: 0042-0000-US; GTIN: 00813502010022; Serial Numbers: IC1140. 5. Product Code: 0042-0010-EU; GTIN: 00813502010961; Serial Numbers: IC5403, IC5404, IC5635, IC5636, IC5666, IC5667, IC5801, IC5904, IC5918, IC5921, IC6421, IC6422, IC8683. 6. Product Code: 0042-0010-UK; GTIN: 00813502010978; Serial Numbers: IC8101, IC8102, IC8103, IC8104, IC8544, IC9446, IC9449, IC11440. 7. Product Code: 0042-0040-CA; GTIN: 00813502012958; Serial Numbers: IC10306, IC10312, IC12604, IC12891, IC12895, IC12892, IC12605, IC12894. 8. Product Code: 0042-0040-EU; GTIN: 0081350201... [TRUNCATED] |
| Postal code | 01923-2575 |
| Report date | 20260408 |
| Product type | Devices |
| Product quantity | 62 units |
| Reason for recall | Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260216 |
| Initial firm notification | |
| Center classification date | 20260327 |
Overview
- Recalling FirmAbiomed, Inc.
- StatusOngoing
- Risk LevelClass I
- DistributionInternational distribution in the country of Australia, Canada, France, Germany, Kuwait.