Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt suppor...
Reason for Recall / Hazard
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
Product Description & Identification
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Affected Products
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Additional Source Details
| Field | Value |
|---|---|
| City | Ellis |
| State | KS |
| Event id | 98332 |
| Address 1 | 204 W 2nd St |
| Address 2 | N/A |
| Code info | UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025 |
| Postal code | 67637-2026 |
| Report date | 20260304 |
| Product type | Devices |
| Product quantity | 4286 |
| Reason for recall | Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260121 |
| Initial firm notification | |
| Center classification date | 20260226 |
Overview
- Recalling FirmAgiliti Health - Ellis
- StatusOngoing
- Risk LevelClass II
- DistributionUS Nationwide distribution.