Home/Recalls/AU-8a0612cf-cfec-42f3-a4a1-339f2a0cab38
Australia ProductsPharmacy and me...

Accu-Chek Solo Cannula Assembly & Pump Holder and Accu-Chek Solo Supply Kit

Published: December 23, 2022Recall ID: 8a0612cf-cfec-42f3-a4a1-339f2a0cab38Category: generalCountry: Australia

Reason for Recall / Hazard

Insulin leakage has occurred with the Accu-Chek Solo Cannula Assembly and Pump Holder. If the insulin leakage is not immediately noticed when using the Accu-Chek Solo system this could potentially lead to an under-delivery of insulin.

Product Description & Identification

Reason / Defects: Insulin leakage has occurred with the Accu-Chek Solo Cannula Assembly and Pump Holder. If the insulin leakage is not immediately noticed when using the Accu-Chek Solo system this could potentially lead to an under-delivery of insulin. Hazards: If insulin is under-delivered, this may lead to adverse health events including severe hyperglycaemia or the user may experience diabetic ketoacidosis (DKA) and may require hospitalisation or intervention from a medical professional. What to do: Consumers should check whether they have an affected product and if so stop using them. Consumers should contact Roche Diabetes Customer Care to make arrangements to receive a replacement. For more details see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01541-1 For further information, please contact Roche Diabetes Customer Care by phone on 1800 251 516.

Affected Products

Accu-Chek Solo Cannula Assembly & Pump Holder (9 mm; 10+3 & 6 mm; 10+3) and Accu-Chek Solo Supply Kit (9 mm) & (6 mm) Cannula lot numbers: 847468303 and lower Reference numbers: 07858833001, 07858841001, 07858817001 & 07858825001 ARTG 308704 (Roche Diabetes Care Australia Pty Ltd - Insulin infusion patch pump administration set) ARTG 311432 (Roche Diabetes Care Australia Pty Ltd - Ambulatory insulin infusion administration set/reservoir kit)

Additional Source Details

FieldValue
Guid › #text8a0612cf-cfec-42f3-a4a1-339f2a0cab38
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/accu-chek-solo-cannula-assembly-pump-holder-and-accu-chek-solo-supply-kit
Pub DateFri, 23 Dec 2022 00:00:00 +0000

Overview

  • Risk LevelPharmacy and medical products
  • DistributionAustralia
Official Agency Alert