Inicio/Ver Todos los Retiros/AU-da73e5e5-17ee-4127-80f7-559cec38797d
Australia ProductsPharmacy and me...

Alpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump

Publicado: 8 de enero de 2021ID de Retiro: da73e5e5-17ee-4127-80f7-559cec38797dCategoría: Productos de ConsumoPaís: Australia
Australia
Safety AlertAustralia ProductsAustraliaAustralia

Alpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump

🌐
Traducir AvisoTranslate notice text, hazard warnings, and descriptions into any supported language.

Motivo del Retiro / Peligro

The mains cables may become damaged through incorrect stowage or misuse.

Descripción del Producto e Identificación

Reason / Defects: The mains cables may become damaged through incorrect stowage or misuse. Hazards: If the mains cables have been damaged, this can result in electric shock, which can cause serious injury or death. What to do: Arjo Australia will contact affected consumers in writing. This issue will be rectified by increasing the cable wall thickness and incorporating an improved grommet. Consumers should: 1. Verify whether they possess any Alpha Active 3, Alpha Active 4 or Alpha Active 5 pumps and ensure that all caregivers and users are made aware of this issue. 2. Make sure the cable is disconnected from the power socket and then check carefully all electrical connections for signs of damage or excessive wear. If any damage is observed, remove the device from use and contact Arjo Australia. 3. The device can remain in use as long as the cable is not frayed or worn and is secured properly to prevent being caught or entrapped in the bed mechanism as advised in the Instruction for Use. For further information, consumers can contact Arjo Australia on 1800 072 040.

Productos Afectados

Alpha Active 3, Alpha Active 4 and Alpha Active 5 medical pumps ARTG 288608 (Arjo Australia Pty Ltd - Alternating-pressure bed mattress overlay system)

Additional Source Details

FieldValue
Guid › #textda73e5e5-17ee-4127-80f7-559cec38797d
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/alpha-active-3-alpha-active-4-and-alpha-active-5-medical-pump
Pub DateFri, 08 Jan 2021 00:00:00 +0000

Consumer Action Protocol: What You Should Do Immediately

Safety Steps
1Verify Product Identifiers

Check the model number, serial number, and date stamp on the rating label on the back or underside of the product.

2Isolate & Discontinue Use

Immediately discontinue using the product. Unplug electrical items and keep juvenile items out of reach of children.

3Claim Free Statutory Remedy

Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.

4Report Adverse Incidents

If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (Australia Products).

Your Legal Rights as a Consumer

Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.

Frequently Asked Questions About This Recall

FAQ Guide

Q1:What should I do immediately if I own or bought Alpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump?

Immediately stop using the product and keep it out of reach of children. Disconnect power if electrical, and do not attempt makeshift home repairs.

Q2:Can I get a full refund or free repair for this recall?

Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.

Q3:Do I need my original sales receipt to claim a remedy or refund?

In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.

Q4:Where can I verify official information for this safety alert?

This safety alert is aggregated directly from official regulatory agency records (Australia Products). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.

Revise otros productos en su hogar

Herramientas Gratuitas

Safety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:

Resumen

  • Nivel de RiesgoPharmacy and medical products
  • DistributionAustralia
Recall Hazard Severity
Caution (Class III)
Alerta Oficial de la Agencia

📢 Compartir advertencia

Alert family and friends to keep them safe. Copy this quick alert message:

🚨 SAFETY ALERT: Alpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump

Check if your household is affected:
https://therecallfile.com/es/recalls/alpha-active-3-alpha-active-4-and-alpha-active-5-medical-pump/AU-da73e5e5-17ee-4127-80f7-559cec38797d
WhatsAppFacebook

Unirse a la lista de alertas

Reciba alertas por correo electrónico cuando se publiquen retiros críticos.

Support The Recall File

Help keep official recall alerts free, accessible, and fast for everyone.

Donate with PayPal

Supports USD, EUR, GBP, International Cards & PayPal

Spread the WordAlpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump

Install The Recall File App

Check safety warnings, scan your pantry, and monitor your vehicles directly from your home screen.