Dispositivos Médicos Recalls
Defective surgical instruments, diagnostic equipment, and health device recalls.
Cobalt-chrome modular neck hip replacements: risk of metal-wear effects and revision surgery (DSI/2026/010)
Cobalt-chrome modular neck hip replacements: risk of metal-wear effects and revision surgery (DSI/2026/010)
An MHRA investigation has found increased risks of wear and corrosion associated with multiple cobalt-chrome (Co-Cr) modular neck hip stem components, including a potential increased occurrence of rev...
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
NGPod pH testing device: quarantine and dispose of all devices (DSI/2026/009)
NGPod pH testing device: quarantine and dispose of all devices (DSI/2026/009)
The NGPod handheld device and NGPod sensor, used to test pH to aid the placement of nasogastric tubes, is being withdrawn from the UK market. The manufacturer has ceased trading in the UK. Users shoul...
Field Safety Notices: 17 to 21 August 2026
Field Safety Notices: 17 to 21 August 2026
List of Field Safety Notices from 17 to 21 August 2026....
60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 Intravascular Administration Set and Extension Set used to administer fluids.
60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 Intravascular Administration Set and Extension Set used to administer fluids.
Products may not have received adequate sterilization processing.
Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16
Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16
The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.
GM Helix Implant, Endosseous Dental Implant Article No. 109.984
GM Helix Implant, Endosseous Dental Implant Article No. 109.984
There may be a mismatch between the implant contained inside the packaging and the information described on the outer carton label. If the the incorrect implant is installed, the issue could cause bone compression or difficulty during installation in situations where there is sufficient bone volume
Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit.
Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit.
Due to off-the-shelf display port component that has audio/video failures
Cardinal Health Presource Kits: 1) SEPTO RHINO PACK , Catalog Number: SEN73NSKCF; 2) PACK NEURO MINOR CRANIOTOMY , Catalog Number: SNE30CNOHE; 3) NEURO MAJOR 2 PACK , Catalog Number: SNE30CRJGY; 4) NEURO PACK , Catalog Number: SNE43NEE13; 5) LAMINECTOMY PACK , Catalog Number: SNE43NEE13; 7) LAMINECTOMY PACK , Catalog Number: SNE5CLPB13; 8) PEDI SPINE R , Catalog Number: SNE69PSKR4; 9) SHUNT PACK , Catalog Number: SNE7CSHY15; 10) SHUNT PACK , Catalog Number: SNE7CSHY16; 11) ENS ORTHO FUSION POSTERIEURE 3, Catalog Number: SOP30ODSCL
Cardinal Health Presource Kits: 1) SEPTO RHINO PACK , Catalog Number: SEN73NSKCF; 2) PACK NEURO MINOR CRANIOTOMY , Catalog Number: SNE30CNOHE; 3) NEURO MAJOR 2 PACK , Catalog Number: SNE30CRJGY; 4) NEURO PACK , Catalog Number: SNE43NEE13; 5) LAMINECTOMY PACK , Catalog Number: SNE43NEE13; 7) LAMINECTOMY PACK , Catalog Number: SNE5CLPB13; 8) PEDI SPINE R , Catalog Number: SNE69PSKR4; 9) SHUNT PACK , Catalog Number: SNE7CSHY15; 10) SHUNT PACK , Catalog Number: SNE7CSHY16; 11) ENS ORTHO FUSION POSTERIEURE 3, Catalog Number: SOP30ODSCL
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form
Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form
Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim may inadvertently contain implants that are 12mm in length instead of 14mm as the labeling would suggest.
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software Version Numbers: R3.0 and R3.1; Interventional fluoroscopic x-ray system
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software Version Numbers: R3.0 and R3.1; Interventional fluoroscopic x-ray system
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
Cardinal Health Presource Kits: 1) NEWNAN SMR PACK, Catalog Number: SEN21SPPMB
Cardinal Health Presource Kits: 1) NEWNAN SMR PACK, Catalog Number: SEN21SPPMB
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.
ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.
The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.
Medline Micro-Kill Bleach Germicidal Bleach Wipes,7" x 8", 150 count, 150 count, Product Number MSC351400AN
Medline Micro-Kill Bleach Germicidal Bleach Wipes,7" x 8", 150 count, 150 count, Product Number MSC351400AN
Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.
GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.
GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1006. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.
10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).
Boston Scientific Enhance Transcarotid/Peripheral Access Kit REF: SR-4F21G7D-MP (Kit-075-03) Sheath with Dilator , Stiffen Dilator, Nitinol Mandrel Wire w/ Radiopaque Tip [Au}, Needle, Extension Tube with Stopcock
Boston Scientific Enhance Transcarotid/Peripheral Access Kit REF: SR-4F21G7D-MP (Kit-075-03) Sheath with Dilator , Stiffen Dilator, Nitinol Mandrel Wire w/ Radiopaque Tip [Au}, Needle, Extension Tube with Stopcock
Black depth markings may detach from the outer sheath of the coaxial dilator introducer component of transcarotid-peripheral access kits.
Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX
Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX
Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
Cardinal Health Presource Kits: 1) FAT GRAFT ADD KIT , Catalog Number: 21-4720B; 2) TRAY FAT GRAFT ADD , Catalog Number: 21-4720C; 3) KIT,BOSC MICROLARYNGOSCOPY , Catalog Number: PECGMLBOB; 4) KIT,ASLMC NASAL PACK , Catalog Number: PECGNAAU8; 5) KIT,ASLMC NASAL PACK , Catalog Number: PECGNAAU8; 6) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLH9; 7) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLHC; 8) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLHD; 9) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLHE; 10) KIT,NEURO , Catalog Number: PN21OTMC6; 11) KIT,NEURO , Catalog Number: PN21OTMC7; 12) KIT,NEURO , Catalog Number: PN21OTMCC; 13) KIT,NEURO , Catalog Number: PN21OTMCD; 14) KIT, MER CRANI , Catalog Number: PN33CRMYB; 15) KIT,CHP BASIC NEURO , Catalog Number: PN36NBCHB; 16) KIT, CRANIOTOMY , Catalog Number: PN40CKC27; 17) KIT, CRANIOTOMY , Catalog Number: PN40CKC28; 18) KIT, CRANIOTOMY , Catalog Number: PN40CKC29; 19) KIT, CRANIOTOMY , Catalog Number: PN40CKCFC; 20) KIT, BAL LUMBAR LAMINECTOMY , Catalog Number: PNCGLLBMB; 21) KIT, BAL LUMBAR LAMINECTOMY , Catalog Number: PNCGLLBMC; 22) BASIC PACK , Catalog Number: SBA73BPGCA; 23) FACIAL FEMINIZATION PACK , Catalog Number: SBA73FFKW9; 24) E BASIC MILLER PACK , Catalog Number: SBA95BMHH6; 25) E BASIC MILLER PACK , Catalog Number: SBA95BMHH7; 26) E BASIC MILLER PACK , Catalog Number: SBA95BMHH8; 27) E BASIC MILLER PACK , Catalog Number: SBA95BMHHF; 28) E BASIC MILLER PACK , Catalog Number: SBA95BMHHG; 29) E BASIC MILLER PACK , Catalog Number: SBA95BMHHH; 30) TRAM PACK , Catalog Number: SBA95TPD10; 31) TRAM PACK , Catalog Number: SBA95TPD11; 32) TRAM PACK , Catalog Number: SBA95TPDU6; 33) TRAM PACK , Catalog Number: SBA95TPDU7; 34) TRAM PACK , Catalog Number: SBA95TPDU8; 35) TRAM PACK , Catalog Number: SBA95TPDU9; 36) TRAM PACK , Catalog Number: SBA95TPDUP; 37) TRAM PACK , Catalog Number: SBA95TPDUQ; 38) PLASTIC SURGERY PACK ,
Cardinal Health Presource Kits: 1) FAT GRAFT ADD KIT , Catalog Number: 21-4720B; 2) TRAY FAT GRAFT ADD , Catalog Number: 21-4720C; 3) KIT,BOSC MICROLARYNGOSCOPY , Catalog Number: PECGMLBOB; 4) KIT,ASLMC NASAL PACK , Catalog Number: PECGNAAU8; 5) KIT,ASLMC NASAL PACK , Catalog Number: PECGNAAU8; 6) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLH9; 7) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLHC; 8) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLHD; 9) KIT, TRIPOINT SPINE , Catalog Number: PN1DTSLHE; 10) KIT,NEURO , Catalog Number: PN21OTMC6; 11) KIT,NEURO , Catalog Number: PN21OTMC7; 12) KIT,NEURO , Catalog Number: PN21OTMCC; 13) KIT,NEURO , Catalog Number: PN21OTMCD; 14) KIT, MER CRANI , Catalog Number: PN33CRMYB; 15) KIT,CHP BASIC NEURO , Catalog Number: PN36NBCHB; 16) KIT, CRANIOTOMY , Catalog Number: PN40CKC27; 17) KIT, CRANIOTOMY , Catalog Number: PN40CKC28; 18) KIT, CRANIOTOMY , Catalog Number: PN40CKC29; 19) KIT, CRANIOTOMY , Catalog Number: PN40CKCFC; 20) KIT, BAL LUMBAR LAMINECTOMY , Catalog Number: PNCGLLBMB; 21) KIT, BAL LUMBAR LAMINECTOMY , Catalog Number: PNCGLLBMC; 22) BASIC PACK , Catalog Number: SBA73BPGCA; 23) FACIAL FEMINIZATION PACK , Catalog Number: SBA73FFKW9; 24) E BASIC MILLER PACK , Catalog Number: SBA95BMHH6; 25) E BASIC MILLER PACK , Catalog Number: SBA95BMHH7; 26) E BASIC MILLER PACK , Catalog Number: SBA95BMHH8; 27) E BASIC MILLER PACK , Catalog Number: SBA95BMHHF; 28) E BASIC MILLER PACK , Catalog Number: SBA95BMHHG; 29) E BASIC MILLER PACK , Catalog Number: SBA95BMHHH; 30) TRAM PACK , Catalog Number: SBA95TPD10; 31) TRAM PACK , Catalog Number: SBA95TPD11; 32) TRAM PACK , Catalog Number: SBA95TPDU6; 33) TRAM PACK , Catalog Number: SBA95TPDU7; 34) TRAM PACK , Catalog Number: SBA95TPDU8; 35) TRAM PACK , Catalog Number: SBA95TPDU9; 36) TRAM PACK , Catalog Number: SBA95TPDUP; 37) TRAM PACK , Catalog Number: SBA95TPDUQ; 38) PLASTIC SURGERY PACK ,
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
Medline Micro-Kill Bleach Germicidal Bleach Wipes, 6" x 5", 150 count, Product Number MSC351410AN
Medline Micro-Kill Bleach Germicidal Bleach Wipes, 6" x 5", 150 count, Product Number MSC351410AN
Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.
Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics
Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics
Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0 CLR 15CM P12X12 VLOCM0003 VLOCM0013 V-LOC* 90 4-0 CLR 30CM P12X12 VLOCM0013 VLOCM0024 V-LOC* 90 3-0 CLR 45CM P12X12 VLOCM0024 VLOCM0123 V-LOC* 90 4-0 CLR 45CM P14X12 VLOCM0123 VLOCM0315 V-LOC* 90 2-0 VIO 30CM GS21 VLOCM0315 VLOCM0603 V-LOC* 90 4-0 VIO 15CM V20X12 VLOCM0603 VLOCM0604 V-LOC* 90 3-0 VIO 15CM V20X12 VLOCM0604 VLOCM1203 V-LOC* 90 4-0 VIO 15CM CV15 VLOCM1203 VLOCM2115 V-LOC* 90 2-0 VIO 30CM GS22 VLOCM2115
V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0 CLR 15CM P12X12 VLOCM0003 VLOCM0013 V-LOC* 90 4-0 CLR 30CM P12X12 VLOCM0013 VLOCM0024 V-LOC* 90 3-0 CLR 45CM P12X12 VLOCM0024 VLOCM0123 V-LOC* 90 4-0 CLR 45CM P14X12 VLOCM0123 VLOCM0315 V-LOC* 90 2-0 VIO 30CM GS21 VLOCM0315 VLOCM0603 V-LOC* 90 4-0 VIO 15CM V20X12 VLOCM0603 VLOCM0604 V-LOC* 90 3-0 VIO 15CM V20X12 VLOCM0604 VLOCM1203 V-LOC* 90 4-0 VIO 15CM CV15 VLOCM1203 VLOCM2115 V-LOC* 90 2-0 VIO 30CM GS22 VLOCM2115
Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection, 0.9%, USP 10mL Ampule Model/Catalog Number: AMPUL10MLK. Sodium Chloride Flush for IV tubing systems and indwelling intravascular access devices.
Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection, 0.9%, USP 10mL Ampule Model/Catalog Number: AMPUL10MLK. Sodium Chloride Flush for IV tubing systems and indwelling intravascular access devices.
Manufacturing firm, Huons, was cited for issues regarding product quality and sterility assurance. If the product sterility is compromised, patients may be exposed to microorganisms, endotoxins, or foreign matter. This may trigger inflammatory responses, infections, micro-clots, and organ failure.
navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number 09453733001 UDI-DI Code: 761333602095AQ In the United States, navify Digital Pathology software (previously known as uPath Enterprise Software) is a component of the Roche Digital Pathology Dx device cleared by the US FDA for Whole Slide Imaging purposes. Roche Digital Pathology Dx is an automated digital slide creation, viewing, and management system.
navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number 09453733001 UDI-DI Code: 761333602095AQ In the United States, navify Digital Pathology software (previously known as uPath Enterprise Software) is a component of the Roche Digital Pathology Dx device cleared by the US FDA for Whole Slide Imaging purposes. Roche Digital Pathology Dx is an automated digital slide creation, viewing, and management system.
Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window