Home/Recalls/CPSC-12742
CPSCGeneral Alert

TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard

Published: July 18, 2012Recall ID: 12742Category: generalCountry: US

Reason for Recall / Hazard

The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.

Product Description & Identification

The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.

Affected Products

Ondal AC2000 (PDI) Television Mounting System

Additional Source Details

FieldValue
U R Lhttps://www.cpsc.gov/Recalls/2012/TRUMPF-Medical-Systems-Recalls-Television-Mounting-System-Due-to-Falling-Hazard
TitleTRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard
Images › U R Lhttps://www.cpsc.gov/s3fs-public/Recall.2012.12742.12742.jpg
Images › CaptionTRUMPF Medical Systems television mounting system
Hazards › NameThe arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Injuries › NameThe firm received one report of a mounting system falling while being positioned. No injuries have been reported.
Products › NameOndal AC2000 (PDI) Television Mounting System
Products › TypeTelevisions, Mounts, Guards & Antennas
Products › Category I D2169
Products › Number Of Units215 distributed to 5 hospitals
Recall I D1200
Remedies › NameSecure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Importers › 1 › NameOndal Medical Systems of America Inc., of Sandston, VA
Importers › 2 › NameTRUMPF Medical Systems Inc., of Charleston, SC
Retailers › NameTRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.
Recall Date2012-07-18T00:00:00
DescriptionThe mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.
Distributors › NameTRUMPF Medical Systems Inc., of Charleston, SC
Recall Number12742
Remedy Options › OptionRepair
Consumer ContactFor additional information, contact TRUMPF Medical Systems Inc. toll-free at (888) 474-9359 between 8 a.m. and 5 p.m. ET Monday through Friday, or email Lindsey.ronnenberg@us.trumpf.com
Last Publish Date2016-04-21T00:00:00
Manufacturer Countries › CountryGermany