Home/Recalls/FDA-Z-2714-2026
FDA DevicesClass II

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Published: July 15, 2026Recall ID: Z-2714-2026Category: devicesCountry: US
US
Medical DeviceFDA DevicesUSUS

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Recalling Firm: Integra LifeSciences Corp. (NeuroSciences)

Reason for Recall / Hazard

The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

Product Description & Identification

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Additional Source Details

FieldValue
CityPrinceton
StateNJ
Event id99193
Address 11100 Campus Rd
Address 2N/A
Code infoProduct Code 3032, UDI 10381780268468; Product Code 3055, UDI 10381780268505; Product Code 3069, UDI 10381780268512; Product Code 3076, UDI 10381780268529; Product Code 3086, UDI 10381780268536; Product Code 3092, UDI 10381780268550; Product Code 3093, UDI 10381780268567; Product Code 3096, UDI 10381780268581; Product Code 3098, UDI 10381780268598; Product Code 3099, UDI 10381780268604; Product Code 3133, UDI 10381780268680; Product Code 3137, UDI 10381780268697; Product Code 3142, UDI 10381780268710; Product Code 3146, UDI 10381780268734; Product Code 3148, UDI 10381780268758; Product Code 3176, UDI 10381780268826; Product Code 3177, UDI 10381780268833; Product Code 3185, UDI 10381780268857; Product Code 3201, UDI 10381780268871; Product Code 3203, UDI 10381780268888; Product Code 3216, UDI 10381780268895; Product Code 3217, UDI 10381780268901; Product Code 3218, UDI 10381780268918; Product Code 3219, UDI 10381780268925; Product Code 3220, UDI 10381780268932; ... [TRUNCATED]
Postal code08540-6650
Report date20260715
Product typeDevices
Product quantity633 units
Reason for recallThe current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260605
Initial firm notificationLetter
Center classification date20260709

Check Other Products In Your Household

Free Safety Tools

Safety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:

Overview

  • Recalling FirmIntegra LifeSciences Corp. (NeuroSciences)
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brunei Darussalam, Bulgaria, Canada, Chile, China, Czechia, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea (the Republic of), Kuwait, Lithuania, Malaysia, Malta, Mauritius, Mexico, Nepal, Netherlands (the), New Zealand, Norway, Pakistan, Palestine, State of, Philippines (the), Poland, Portugal, Puerto Rico, Qatar, Republic of North Macedonia, R¿union, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan (Province of China), Thailand, Trinidad and Tobago, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland, (the), Viet Nam.
Recall Hazard Severity
High Alert (Class I)
Official Agency Alert

📢 Share Warning

Alert family and friends to keep them safe. Copy this quick alert message:

🚨 SAFETY ALERT: Integra LifeSciences Corp. (NeuroSciences) - OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Check if your household is affected:
https://therecallfile.com/recalls/omnitract-surgical-retractor-system-which-includes-633-of-the-firm-s-p/FDA-Z-2714-2026
WhatsAppFacebook

Join Alertlist

Get instant email updates whenever critical safety recalls like this are published.

Featured on Product Hunt

Help support our public safety mission and check out our community launch.

The Recall File - The Google for Product Recalls | Product Hunt

Support The Recall File

Help keep official recall alerts free, accessible, and fast for everyone.

Donate with PayPal

Supports USD, EUR, GBP, International Cards & PayPal

Spread the WordOmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Install The Recall File App

Check safety warnings, scan your pantry, and monitor your vehicles directly from your home screen.