Home/Recalls/FDA-Z-1319-2026
FDA DevicesClass I

Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Elec...

Published: February 18, 2026Recall ID: Z-1319-2026Category: devicesCountry: US

Reason for Recall / Hazard

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Product Description & Identification

Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;

Affected Products

Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;

Additional Source Details

FieldValue
CityNorthfield
StateIL
Event id98277
Address 13 Lakes Dr
Address 2N/A
Code infoBD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP250808; BD710FJ282RRH UDI-DI 10197344044054 Lots EP250429 EP250808; BD710FJ282CRH UDI-DI (case) 40197344044048 (ea) 10197344044047 Lots EP250429 EP250808
Postal code60093-2753
Report date20260218
Product typeDevices
Product quantity228
Reason for recallThese lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251222
Initial firm notificationLetter
Center classification date20260206

Overview

  • Recalling FirmMedline Industries, LP
  • StatusOngoing
  • Risk LevelClass I
  • DistributionDistribution US nationwide.
Official Agency Alert