Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Elec...
Reason for Recall / Hazard
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Product Description & Identification
Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;
Affected Products
Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;
Additional Source Details
| Field | Value |
|---|---|
| City | Northfield |
| State | IL |
| Event id | 98277 |
| Address 1 | 3 Lakes Dr |
| Address 2 | N/A |
| Code info | BD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP250808; BD710FJ282RRH UDI-DI 10197344044054 Lots EP250429 EP250808; BD710FJ282CRH UDI-DI (case) 40197344044048 (ea) 10197344044047 Lots EP250429 EP250808 |
| Postal code | 60093-2753 |
| Report date | 20260218 |
| Product type | Devices |
| Product quantity | 228 |
| Reason for recall | These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025 |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20251222 |
| Initial firm notification | Letter |
| Center classification date | 20260206 |
Overview
- Recalling FirmMedline Industries, LP
- StatusOngoing
- Risk LevelClass I
- DistributionDistribution US nationwide.