FDA DevicesClass II
LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUD...
Published: May 13, 2026Recall ID: Z-2049-2026Category: devicesCountry: US
Reason for Recall / Hazard
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
Product Description & Identification
LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5
Additional Source Details
| Field | Value |
|---|---|
| City | Saint Paul |
| State | MN |
| Event id | 98679 |
| Address 1 | 4100 Hamline Ave N |
| Code info | LUX-Dx II Plus UDI-DI: 00802526620706 (US), 00802526623707 (EU/AU); SERVER SW LATITUDE DRAGON EU UDI-DI: 00802526613821; SERVER SW LATITUDE DRAGON AU UDI-DI: 00802526613814; Serial Number: 100085 100676 100909 100944 101811 102341 102405 102864 102956 102978 103018 105283 105697 106372 106429 106690 107386 107460 108132 108429 108433 108641 108697 108805 108861 108901 109663 109699 109836 110700 110737 110858 111048 111119 111269 111290 111362 111780 111987 112265 112549 113400 113495 113720 114072 114393 115310 115411 115835 115862 115890 115981 116066 116090 116197 117180 117533 117560 117820 117924 117960 119462 120931 121080 121203 121267 121483 122855 123305 123412 124378 124458 124557 124791 124987 125008 126249 127182 127441 127758 127821 128106 128506 129528 129763 129782 130474 131469 132487 132549 133195 133786 133989 134600 135362 136038 136746 136884 137047 137134 137319 137449 138039 138059 138334 138759 139080 139165 140319 141181 142273 143441 143887 144590 144777 145284... [TRUNCATED] |
| Postal code | 55112-5700 |
| Report date | 20260513 |
| Product type | Devices |
| Product quantity | 148 units |
| Reason for recall | For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260330 |
| Initial firm notification | Letter |
| Center classification date | 20260501 |
Overview
- Recalling FirmBoston Scientific Corporation
- StatusOngoing
- Risk LevelClass II
- DistributionWorldwide distribution - US Nationwide and the country of Belgium.