Home/Recalls/FDA-Z-1588-2026
FDA DevicesClass I

Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Nin...

Published: April 1, 2026Recall ID: Z-1588-2026Category: devicesCountry: US

Reason for Recall / Hazard

If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.

Product Description & Identification

Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd

Additional Source Details

FieldValue
CityDenver
StateCO
Event id98453
Address 15330 Fox St
Code infoUPC: 0-24291-25607-6, UDI-DI: 00024291256076
Postal code80216-1630
Report date20260401
Product typeDevices
Product quantity28,457
Reason for recallIf heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260218
Initial firm notificationLetter
Center classification date20260320

Overview

  • Recalling FirmNavajo Manufacturing Company
  • StatusOngoing
  • Risk LevelClass I
  • DistributionUS Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.