FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Published: February 11, 2026Recall ID: Z-1299-2026Category: devicesCountry: US
The Recall File - The Google for Product Recalls | Product Hunt
US
Medical DeviceFDA DevicesUSUS

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Recalling Firm: FUJIFILM Healthcare Americas Corporation

Reason for Recall / Hazard

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Product Description & Identification

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Affected Products

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Additional Source Details

FieldValue
CityLexington
StateMA
Event id98124
Address 181 Hartwell Ave Ste 300
Address 2N/A
Code infoModel Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
Postal code02421-3160
Report date20260211
Product typeDevices
Product quantity16 units
Reason for recallIt was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260109
Initial firm notificationLetter
Center classification date20260205

Check Other Products In Your Household

Free Safety Tools

Safety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:

Overview

  • Recalling FirmFUJIFILM Healthcare Americas Corporation
  • StatusOngoing
  • Risk LevelClass II
  • DistributionUS Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Recall Hazard Severity
High Alert (Class I)
Official Agency Alert

📢 Share Warning

Alert family and friends to keep them safe. Copy this quick alert message:

🚨 SAFETY ALERT: FUJIFILM Healthcare Americas Corporation - FDR Visionary Suite;  Model Number: CH-200;  Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Check if your household is affected:
https://therecallfile.com/recalls/fdr-visionary-suite-model-number-ch-200-version-1-566-16130-23-2-566-1/FDA-Z-1299-2026
WhatsAppFacebook

Join Alertlist

Get instant email updates whenever critical safety recalls like this are published.

Spread the WordFDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Install The Recall File App

Check safety warnings, scan your pantry, and monitor your vehicles directly from your home screen.