FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Recalling Firm: FUJIFILM Healthcare Americas Corporation
Reason for Recall / Hazard
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Product Description & Identification
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Affected Products
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Additional Source Details
| Field | Value |
|---|---|
| City | Lexington |
| State | MA |
| Event id | 98124 |
| Address 1 | 81 Hartwell Ave Ste 300 |
| Address 2 | N/A |
| Code info | Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002; |
| Postal code | 02421-3160 |
| Report date | 20260211 |
| Product type | Devices |
| Product quantity | 16 units |
| Reason for recall | It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260109 |
| Initial firm notification | Letter |
| Center classification date | 20260205 |
Check Other Products In Your Household
Free Safety ToolsSafety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:
Overview
- Recalling FirmFUJIFILM Healthcare Americas Corporation
- StatusOngoing
- Risk LevelClass II
- DistributionUS Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
📢 Share Warning
Alert family and friends to keep them safe. Copy this quick alert message:
🚨 SAFETY ALERT: FUJIFILM Healthcare Americas Corporation - FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3) Check if your household is affected: https://therecallfile.com/recalls/fdr-visionary-suite-model-number-ch-200-version-1-566-16130-23-2-566-1/FDA-Z-1299-2026