Home/Recalls/FDA-Z-1573-2026
FDA DevicesClass II

Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS...

Published: March 25, 2026Recall ID: Z-1573-2026Category: devicesCountry: US

Reason for Recall / Hazard

Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.

Product Description & Identification

Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package

Affected Products

Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package

Additional Source Details

FieldValue
CityDeerfield
StateIL
Event id98439
Address 11 Baxter Pkwy
Address 2N/A
Code infoLot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table
Postal code60015-4625
Report date20260325
Product typeDevices
Product quantity59
Reason for recallDisplay screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260218
Initial firm notificationLetter
Center classification date20260318

Overview

  • Recalling FirmBaxter Healthcare Corporation
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
Official Agency Alert