Home/Recalls/FDA-Z-1700-2026
FDA DevicesClass II

Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci ...

Published: April 8, 2026Recall ID: Z-1700-2026Category: devicesCountry: US

Reason for Recall / Hazard

Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments

Product Description & Identification

Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/Catalog Number: 420207 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A

Affected Products

Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/Catalog Number: 420207 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A

Additional Source Details

FieldValue
CitySunnyvale
StateCA
Event id98408
Address 11266 Kifer Rd
Code infoLot Code: 420207: Tenaculum Forceps Instrument GTIN: 00886874111659 FDA Medical Device Number: D088920 Material Number: 420207-06 da Vinci System Numbers: SH1378 SH0935 SH0935 SH0935 USG294 SH0761 SH0729 SH1182 SH1378 USG022 SH1378 SH1368 SH0059 SH1015 SH1182 SH0825 USG294 SH1390 USG022 SH0495 SH0495 SH1076 SH1390 SH1390 SH1390 SH1276 SH1385 USG022 SH0547 SH1385 SH1391 SH2144 SH0059 SH1279 SH0059 SH1390 SH1570 SH0059 SH1391 SG704 SH1570 SH0729 SH1385 SH1378 USG518 SH1385 SH1837 USG294 SH1368 SH1390 SH1385 SH1368 SG2026 SH1557 SH2144 SH1385 USG518 SH1279 USG294 SH1391 Material Number: 420207-07 da Vinci System Numbers: SH1391 SH1385 SH1385 SH1279 SH2125 SH0729 SH1368 USG721 SH1385 SH1997 SH2125 USG294 SH1378 SH1997 SH0035 SG704 SH1385 SH1279 SH1279 SH2125 SH1378 SH1378 SH0495 SH1390 SH1390 SH1385 SH2125 SH1378 SH1385 SH1279 USG721 SH1279 SH1837 USG518 SH1385 SH1378 USG022 SH1378 SH1570 SH1385 SH1385 SH0935 SH0935 SG704 SG704 SH1378 USG022 SH1385 SH1279 USG721 SH1279 USG518 SH1837 SH12... [TRUNCATED]
Postal code94086-5304
Report date20260408
Product typeDevices
Product quantity2095 instruments
Reason for recallDue to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251209
Initial firm notificationLetter
Center classification date20260330

Overview

  • Recalling FirmIntuitive Surgical, Inc.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionU.S. Nationwide distribution in the states of AR, CA, CT, FL, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, NC, OH, PA, TX, VA, and WA. The countries of Argentina, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Official Agency Alert