Home/Recalls/FDA-Z-1092-2026
FDA DevicesClass II

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: ...

Published: January 21, 2026Recall ID: Z-1092-2026Category: devicesCountry: US

Reason for Recall / Hazard

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

Product Description & Identification

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Affected Products

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Additional Source Details

FieldValue
CityUppsala
Event id97872
Address 1Bredgrand 18
Code infoLot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module
Postal codeN/A
Report date20260121
Product typeDevices
Product quantity189
Reason for recallDue to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251009
Initial firm notificationLetter
Center classification date20260114

Overview

  • Recalling FirmC-RAD POSITIONING AB
  • StatusOngoing
  • Risk LevelClass II
  • DistributionThe software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
Official Agency Alert