ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
Recalling Firm: Abiomed, Inc.
Reason for Recall / Hazard
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Product Description & Identification
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
Affected Products
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
Additional Source Details
| Field | Value |
|---|---|
| City | Danvers |
| State | MA |
| Event id | 98893 |
| Address 1 | 22 Cherry Hill Dr |
| Code info | Pump Set Code 0048-0003; Introducer Product Code 0052-3046. GTIN: 00813502012279, 00813502011876. All current batches in field. |
| Postal code | 01923-2575 |
| Report date | 20260722 |
| Product type | Devices |
| Product quantity | 87,277 units |
| Reason for recall | A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260611 |
| Initial firm notification | Letter |
| Center classification date | 20260710 |
Check Other Products In Your Household
Free Safety ToolsSafety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:
Overview
- Recalling FirmAbiomed, Inc.
- StatusOngoing
- Risk LevelClass I
- DistributionWorldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).
📢 Share Warning
Alert family and friends to keep them safe. Copy this quick alert message:
🚨 SAFETY ALERT: Abiomed, Inc. - ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories Check if your household is affected: https://therecallfile.com/recalls/abiomed-impella-cp-set-pump-set-code-0048-0003-containing-biomed-impel/FDA-Z-2682-2026
Join Alertlist
Get instant email updates whenever critical safety recalls like this are published.