MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values.

Publié le: 11 février 2026ID du rappel: Z-1238-2026Catégorie: Dispositifs MédicauxPays: États-Unis
US
Medical DeviceFDA DevicesUSUS

MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values.

Recalling Firm: Medtronic MiniMed, Inc.

🌐
Traduire l’alerteTranslate notice text, hazard warnings, and descriptions into any supported language.

Motif du rappel / Danger

A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.

Description du produit & Identification

MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values.

Produits concernés

MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values.

Additional Source Details

FieldValue
CityNorthridge
StateCA
Event id97997
Address 118000 Devonshire St
Address 2N/A
Code infoMiniMed 780G MMT-1884 (US) UDI-DI codes: 00199150003686 00763000414344 00763000545604 00763000578640 00763000586195 00763000640095 00763000758547 00763000819378 00763000819385 00763000819392 00763000865665 00763000924089 00763000924102 00763000924119 76300075854701 76300081937801 76300081938501 76300096001801 76300096003201 76300096004901 Serial Numbers: NG3666779H NG3811803H NG3926926H NG3944480H NG4116699H NG4173817H NG3334724H NG3750453H NG3874105H NG3924178H NG3926632H NG3927673H NG3944520H NG4023801H NG4029254H NG4097850H NG4146571H NG4170515H NG4200310H NG4177489H NG4178626H NG2752574H NG3423286H NG3838934H NG3926282H NG3948645H NG4017463H NG4010052H NG4174745H NG4178395H NG3502087H NG3634564H NG3756694H NG3871741H NG3919620H NG3927825H NG4164005H NG3729206H NG3839883H NG3884748H NG3908444H NG3961353H NG4068853H NG4179772H NG3106884H NG3116129H NG3271492H NG3351525H NG3458389H NG3464645H NG3490545H NG3482017H NG3520991H NG3593959H NG3515336H NG3467707H NG3563980H NG3654085H NG365... [TRUNCATED]
Postal code91325-1219
Report date20260211
Product typeDevices
Product quantity577 units
Reason for recallA software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20251102
Initial firm notificationLetter
Center classification date20260202

Consumer Action Protocol: What You Should Do Immediately

Safety Steps
1Verify Product Identifiers

Inspect the NDC code, batch number, or control number on your medicine bottle or blister packaging to confirm whether your unit is included.

2Isolate & Discontinue Use

Do not abruptly stop critical prescribed medication without physician advice. Contact your dispensing pharmacist immediately for a safe alternative.

3Claim Free Statutory Remedy

Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.

4Report Adverse Incidents

If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (FDA Devices).

Your Legal Rights as a Consumer

Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.

Frequently Asked Questions About This Recall

FAQ Guide

Q1:What should I do immediately if I own or bought MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump...?

Check the lot number and NDC on your prescription label. For essential medications, consult your doctor or dispensing pharmacist immediately before abruptly stopping your dose.

Q2:Can I get a full refund or free repair for this recall?

Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.

Q3:Do I need my original sales receipt to claim a remedy or refund?

In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.

Q4:Where can I verify official information for this safety alert?

This safety alert is aggregated directly from official regulatory agency records (FDA Devices). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.

Vérifiez les autres produits de votre foyer

Outils gratuits

Safety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:

Aperçu

  • Entreprise responsableMedtronic MiniMed, Inc.
  • StatusOngoing
  • Niveau de risqueClass II
  • DistributionWorldwide - U.S. National distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The country of Israel.
Recall Hazard Severity
High Alert (Class I)
Alerte officielle de l’agence

📢 Partager l’alerte

Alert family and friends to keep them safe. Copy this quick alert message:

🚨 SAFETY ALERT: Medtronic MiniMed, Inc. - MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values.

Check if your household is affected:
https://therecallfile.com/fr/recalls/minimed-780g-insulin-pump-catalog-numbers-mmt-1884-mmt-1886-with-softw/FDA-Z-1238-2026
WhatsAppFacebook

Rejoindre la liste d’alerte

Recevez des alertes e-mail instantanées dès qu’un rappel critique est publié.

Support The Recall File

Help keep official recall alerts free, accessible, and fast for everyone.

Donate with PayPal

Supports USD, EUR, GBP, International Cards & PayPal

Spread the WordMiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values.

Install The Recall File App

Check safety warnings, scan your pantry, and monitor your vehicles directly from your home screen.