DxC 700 AU, REF: B86444, B86446
DxC 700 AU, REF: B86444, B86446
Recalling Firm: Beckman Coulter Mishima K.K.
Motif du rappel / Danger
TranslatedA delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.
Description du produit & Identification
TranslatedDxC 700 AU, REF: B86444, B86446
Additional Source Details
| Field | Value |
|---|---|
| City | Sunto-Gun |
| State | N/A |
| Event id | 98346 |
| Address 1 | 454-32 |
| Address 2 | Higashino; Nagaizumi-Cho |
| Code info | REF/UDI-DI/Serial Numbers: Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B86444/14987666542545/2025060063, 2025060073, 2025060074, 2025060075, 2025060079, 2025060085, 2025060086, 2025060087, 2025060091, 2025060098, 2025060099, 2025060124, 2025060125, 2025060126, 2025060127, 2025060128, 2025060129, 2025070139, 2025070156, 2025070159, 2025070160, 2025070161, 2025070164, 2025070167, 2025070168, 2025070169, 2025070178, 2025070180, 2025070181, 2025070182, 2025070184, 2025070186, 2025070191, 2025070192, 2025070193, 2025070194, 2025070204, 2025080208, 2025080217, 2025080219, 2025080236, 2025080237, 2025080238, 2025080241, 2025080242, 2025080246, 2025080247, 2025080250, 2025090266, 2025090267, 2025090270, 2025090271, 2025090273, 2025090274, 2025090275, 2025090281, 2025090283, 2025090285, 2025090286, 2025090287, 2025090289, 2025090290, 2025090291, 2025090292, 2025090293, 2025090294, 2025090295, 2025090296, 2025090297, 2025090298, 2025090299, 202509030... [TRUNCATED] |
| Postal code | N/A |
| Report date | 20260318 |
| Product type | Devices |
| Product quantity | 142 |
| Reason for recall | A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260130 |
| Initial firm notification | Letter |
| Center classification date | 20260311 |
Consumer Action Protocol: What You Should Do Immediately
Safety StepsInspect the NDC code, batch number, or control number on your medicine bottle or blister packaging to confirm whether your unit is included.
Do not abruptly stop critical prescribed medication without physician advice. Contact your dispensing pharmacist immediately for a safe alternative.
Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.
If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (FDA Devices).
Your Legal Rights as a Consumer
Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.
Frequently Asked Questions About This Recall
FAQ GuideQ1:What should I do immediately if I own or bought DxC 700 AU, REF: B86444, B86446?
Check the lot number and NDC on your prescription label. For essential medications, consult your doctor or dispensing pharmacist immediately before abruptly stopping your dose.
Q2:Can I get a full refund or free repair for this recall?
Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.
Q3:Do I need my original sales receipt to claim a remedy or refund?
In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.
Q4:Where can I verify official information for this safety alert?
This safety alert is aggregated directly from official regulatory agency records (FDA Devices). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.
Vérifiez les autres produits de votre foyer
Outils gratuitsSafety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:
Aperçu
- Entreprise responsableBeckman Coulter Mishima K.K.
- StatusOngoing
- Niveau de risqueClass II
- DistributionUS Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.
📢 Partager l’alerte
Alert family and friends to keep them safe. Copy this quick alert message:
🚨 SAFETY ALERT: Beckman Coulter Mishima K.K. - DxC 700 AU, REF: B86444, B86446 Check if your household is affected: https://therecallfile.com/fr/recalls/dxc-700-au-ref-b86444-b86446/FDA-Z-1552-2026
Rejoindre la liste d’alerte
Recevez des alertes e-mail instantanées dès qu’un rappel critique est publié.
Support The Recall File
Help keep official recall alerts free, accessible, and fast for everyone.