Accueil/Tous les Rappels/FDA-Z-2973-2026
FDA DevicesClass II

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Publié le: 26 août 2026ID du rappel: Z-2973-2026Catégorie: Dispositifs MédicauxPays: États-Unis
US
Medical DeviceFDA DevicesUSUS

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Recalling Firm: Beckman Coulter Inc.

🌐
Traduire l’alerteTranslate notice text, hazard warnings, and descriptions into any supported language.

Motif du rappel / Danger

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Description du produit & Identification

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Produits concernés

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Additional Source Details

FieldValue
CityBrea
StateCA
Event id99436
Address 1250 S Kraemer Blvd
Code info1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.
Postal code92821-6232
Report date20260826
Product typeDevices
Reason for recalla software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260715
Initial firm notificationLetter
Center classification date20260817

Consumer Action Protocol: What You Should Do Immediately

Safety Steps
1Verify Product Identifiers

Inspect the NDC code, batch number, or control number on your medicine bottle or blister packaging to confirm whether your unit is included.

2Isolate & Discontinue Use

Do not abruptly stop critical prescribed medication without physician advice. Contact your dispensing pharmacist immediately for a safe alternative.

3Claim Free Statutory Remedy

Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.

4Report Adverse Incidents

If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (FDA Devices).

Your Legal Rights as a Consumer

Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.

Frequently Asked Questions About This Recall

FAQ Guide

Q1:What should I do immediately if I own or bought Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, D...?

Check the lot number and NDC on your prescription label. For essential medications, consult your doctor or dispensing pharmacist immediately before abruptly stopping your dose.

Q2:Can I get a full refund or free repair for this recall?

Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.

Q3:Do I need my original sales receipt to claim a remedy or refund?

In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.

Q4:Where can I verify official information for this safety alert?

This safety alert is aggregated directly from official regulatory agency records (FDA Devices). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.

Vérifiez les autres produits de votre foyer

Outils gratuits

Safety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:

Aperçu

  • Entreprise responsableBeckman Coulter Inc.
  • StatusOngoing
  • Niveau de risqueClass II
  • DistributionWorldwide distribution - US Nationwide.
Recall Hazard Severity
High Alert (Class I)
Alerte officielle de l’agence

📢 Partager l’alerte

Alert family and friends to keep them safe. Copy this quick alert message:

🚨 SAFETY ALERT: Beckman Coulter Inc. - Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Check if your household is affected:
https://therecallfile.com/fr/recalls/beckman-coulter-dxc-500i-clinical-analyzer-modules-using-software-vers/FDA-Z-2973-2026
WhatsAppFacebook

Rejoindre la liste d’alerte

Recevez des alertes e-mail instantanées dès qu’un rappel critique est publié.

Support The Recall File

Help keep official recall alerts free, accessible, and fast for everyone.

Donate with PayPal

Supports USD, EUR, GBP, International Cards & PayPal

Spread the WordBeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Install The Recall File App

Check safety warnings, scan your pantry, and monitor your vehicles directly from your home screen.