Medical DeviceUSFDA Devices

Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777

Cardinal Health 200, LLC
FDA DevicesClass II

Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777

The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes.

Apr 8, 2026Cardinal Health 200, LLC
Medical DeviceUSFDA Devices

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Synthes (USA) Products LLC
FDA DevicesClass II

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

Apr 8, 2026Synthes (USA) Products LLC
Medical DeviceUSFDA Devices

OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem

Oxoid Australia Pty Limited
FDA DevicesClass II

OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem

Oxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to inconclusive result from ambiguous color reactions in wells.

Apr 8, 2026Oxoid Australia Pty Limited
Food & BeverageUSFDA Food

POPPING BOBA RTD MANGO PASSION FRUIT GREEN TEA 12 X 13.8OZ

Pocas International Corp.
FDA FoodClass III

POPPING BOBA RTD MANGO PASSION FRUIT GREEN TEA 12 X 13.8OZ

Packaging integrity issue that may compromise product quality .

Apr 8, 2026Pocas International Corp.
Medical DeviceUSFDA Devices

Paradigm Insulin Pump (MMT-712, MMT-715)

Medtronic MiniMed, Inc.
FDA DevicesClass II

Paradigm Insulin Pump (MMT-712, MMT-715)

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Apr 8, 2026Medtronic MiniMed, Inc.
Medical DeviceUSFDA Devices

Burlington Medical, Caps.

Burlington Medical, LLC
FDA DevicesClass II

Burlington Medical, Caps.

Potential for attenuation degradation over time, decreasing the lifespan.

Apr 8, 2026Burlington Medical, LLC
Medical DeviceUSFDA Devices

Burlington Medical, Thyroid Shield.

Burlington Medical, LLC
FDA DevicesClass II

Burlington Medical, Thyroid Shield.

Potential for attenuation degradation over time, decreasing the lifespan.

Apr 8, 2026Burlington Medical, LLC
Food & BeverageUSFDA Food

POPPING BOBA RTD STRAWBERRY DRAGON F/OOLONG TEA 12 X 13.8OZ

Pocas International Corp.
FDA FoodClass III

POPPING BOBA RTD STRAWBERRY DRAGON F/OOLONG TEA 12 X 13.8OZ

Packaging integrity issue that may compromise product quality .

Apr 8, 2026Pocas International Corp.
Medical DeviceUSFDA Devices

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Abiomed, Inc.
FDA DevicesClass I

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

Apr 8, 2026Abiomed, Inc.
Medical DeviceUSFDA Devices

eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149

Civco Medical Instruments Co. Inc.
FDA DevicesClass II

eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149

There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

Apr 8, 2026Civco Medical Instruments Co. Inc.
Medical DeviceUSFDA Devices

CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc

Mentor Texas, LP.
FDA DevicesClass II

CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc

Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.

Apr 8, 2026Mentor Texas, LP.
Medical DeviceUSFDA Devices

Burlington Medical, Half Aprons.

Burlington Medical, LLC
FDA DevicesClass II

Burlington Medical, Half Aprons.

Potential for attenuation degradation over time, decreasing the lifespan.

Apr 8, 2026Burlington Medical, LLC
Medical DeviceUSFDA Devices

Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT

Medline Industries, LP
FDA DevicesClass II

Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT

Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.

Apr 8, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA

Intuitive Surgical, Inc.
FDA DevicesClass II

Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA

Due to increased complaints for broken/frayed grip cables for reusable instruments.

Apr 8, 2026Intuitive Surgical, Inc.
Medical DeviceUSFDA Devices

Stryker CHROMOPHARE Softlit Ring Surgical Light System REF: CH00000001

Stryker Communications
FDA DevicesClass II

Stryker CHROMOPHARE Softlit Ring Surgical Light System REF: CH00000001

Surgical light assembly may not adequate support the weight of the ceiling cover.

Apr 8, 2026Stryker Communications
Medical DeviceUSFDA Devices

Allura Xper FD20/20 OR Table; System Code: 722039;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD20/20 OR Table; System Code: 722039;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

Apr 8, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Medical DeviceUSFDA Devices

Burlington Medical, Kilt

Burlington Medical, LLC
FDA DevicesClass II

Burlington Medical, Kilt

Potential for attenuation degradation over time, decreasing the lifespan.

Apr 8, 2026Burlington Medical, LLC
Medical DeviceUSFDA Devices

Artoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RH TEXP140RH TEXP150RH Textured, High & Ultra High Profile, Integral Injection Dome, 300cc, 375cc, 535cc, 475cc, 500cc, 600cc, 750cc,

Mentor Texas, LP.
FDA DevicesClass II

Artoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RH TEXP140RH TEXP150RH Textured, High & Ultra High Profile, Integral Injection Dome, 300cc, 375cc, 535cc, 475cc, 500cc, 600cc, 750cc,

Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.

Apr 8, 2026Mentor Texas, LP.
Medical DeviceUSFDA Devices

Sterile Radiology Procedure Kits, Model Number DYNDH1491B

Medline Industries, LP
FDA DevicesClass II

Sterile Radiology Procedure Kits, Model Number DYNDH1491B

There is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility.

Apr 8, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

THORACIC ROBOTS, DYNJ908777B

Medline Industries, LP
FDA DevicesClass II

THORACIC ROBOTS, DYNJ908777B

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Apr 8, 2026Medline Industries, LP
Vehicle RecallUSNHTSA

Bombardier Recreational Products, Inc. - Front Brake Failure

Bombardier Recreational Products, Inc.
NHTSASERVICE BRAKES,...

Bombardier Recreational Products, Inc. - Front Brake Failure

Front brake failure increases the risk of crash or injury.

Apr 7, 2026Bombardier Recreational Products, Inc.
Vehicle RecallUSNHTSA

Hyundai Motor America - Seat Belt Anchors May Detach

Hyundai Motor America
NHTSASEAT BELTS

Hyundai Motor America - Seat Belt Anchors May Detach

A detached seat belt anchor will not adequately restrain the seat occupant, increasing the risk of injury in a crash.

Apr 6, 2026Hyundai Motor America
Food & BeverageUSUSDA FSIS

FSIS Retracts Public Health Alert for Frozen, Ready-to-Eat Chicken Nuggets Due to Updated Laboratory Result

Dorada Foods
USDA FSISPublic Health A...

FSIS Retracts Public Health Alert for Frozen, Ready-to-Eat Chicken Nuggets Due to Updated Laboratory Result

Product Contamination

Apr 6, 2026Dorada Foods
Vehicle RecallUSNHTSA

Greenpower Motor Company Inc. - Warning Label Missing from Cross View Mirror/FMVSS 111

Greenpower Motor Company Inc.
NHTSAEQUIPMENT

Greenpower Motor Company Inc. - Warning Label Missing from Cross View Mirror/FMVSS 111

Without a warning label, the driver may use the altered view to see other vehicles, increasing the risk of a crash.