Inicio/Ver Todos los Retiros/AU-5a23c526-b9a7-43d4-aa8c-66f2215bfff9
Australia ProductsPharmacy and me...

BiPAP A40 Ventilator

Publicado: 30 de agosto de 2022ID de Retiro: 5a23c526-b9a7-43d4-aa8c-66f2215bfff9Categoría: Productos de ConsumoPaís: Australia
🌐
Traducir AvisoTranslate notice text, hazard warnings, and descriptions into any supported language.

Motivo del Retiro / Peligro

Certain devices were built with motor assemblies that may contain non-conforming plastic material.

Descripción del Producto e Identificación

Reason / Defects: Certain devices were built with motor assemblies that may contain non-conforming plastic material. Hazards: If the non-conforming material is present patients may be exposed to: 1. Chemical exposure due to off-gassing, creating a potential biosafety or toxicological hazard. 2. Sudden failure of the device causing a Ventilator Inoperative condition, which may result in asphyxia if not immediately identified and addressed by the care provider. What to do: Philips will replace all the affected devices, users will receive new A40 Pro units. For more details, see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01118-1 If you believe your product is affected and you have not received a letter from Philips, please contact Philips Sleep and Respiratory Care by phone on 1800 009 579.

Productos Afectados

BiPAP A40 Ventilator Serial numbers: V2887246677A0, V288724698F57, V28872472285C, V288724777FF1, V28872479968F, V288724808886, V28872485DF2B, V28872492B24C, V28872495C6F3, V288724981D16, V288725010D13, V2887251114CB and V2887251771FD ARTG 200289 (Philips Electronics Australia Ltd - Positive airway pressure unit, bi-level)

Additional Source Details

FieldValue
Guid › #text5a23c526-b9a7-43d4-aa8c-66f2215bfff9
Guid › @ is Perma Linkfalse
Linkhttps://www.productsafety.gov.au/search-consumer-product-recalls/bipap-a40-ventilator
Pub DateTue, 30 Aug 2022 00:00:00 +0000
Extracted imagehttps://www.productsafety.gov.au/system/files/styles/large/private/Philips%20BiPAP%20A40%20Ventilator.jpg?itok=OYVu0cx2

Consumer Action Protocol: What You Should Do Immediately

Safety Steps
1Verify Product Identifiers

Check the model number, serial number, and date stamp on the rating label on the back or underside of the product.

2Isolate & Discontinue Use

Immediately discontinue using the product. Unplug electrical items and keep juvenile items out of reach of children.

3Claim Free Statutory Remedy

Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.

4Report Adverse Incidents

If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (Australia Products).

Your Legal Rights as a Consumer

Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.

Frequently Asked Questions About This Recall

FAQ Guide

Q1:What should I do immediately if I own or bought BiPAP A40 Ventilator?

Immediately stop using the product and keep it out of reach of children. Disconnect power if electrical, and do not attempt makeshift home repairs.

Q2:Can I get a full refund or free repair for this recall?

Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.

Q3:Do I need my original sales receipt to claim a remedy or refund?

In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.

Q4:Where can I verify official information for this safety alert?

This safety alert is aggregated directly from official regulatory agency records (Australia Products). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.

Revise otros productos en su hogar

Herramientas Gratuitas

Safety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:

Resumen

  • Nivel de RiesgoPharmacy and medical products
  • DistributionAustralia
Recall Hazard Severity
Caution (Class III)
Alerta Oficial de la Agencia

📢 Compartir advertencia

Alert family and friends to keep them safe. Copy this quick alert message:

🚨 SAFETY ALERT: BiPAP A40 Ventilator

Check if your household is affected:
https://therecallfile.com/es/recalls/bipap-a40-ventilator/AU-5a23c526-b9a7-43d4-aa8c-66f2215bfff9
WhatsAppFacebook

Unirse a la lista de alertas

Reciba alertas por correo electrónico cuando se publiquen retiros críticos.

Support The Recall File

Help keep official recall alerts free, accessible, and fast for everyone.

Donate with PayPal

Supports USD, EUR, GBP, International Cards & PayPal

Spread the WordBiPAP A40 Ventilator

Install The Recall File App

Check safety warnings, scan your pantry, and monitor your vehicles directly from your home screen.